Philips Medical Systems (Cleveland) Inc: The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. Product Usage: The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories

Recall number
Z-2245-2012
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. Product Usage: The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories
Status
Terminated
Initiated
2012-08-03
Posted
2012-08-22
Terminated
2015-07-23
Reason
The patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch.
Root cause
Device Design
Action
Philips Medical Systems (Cleveland), Inc. sent an Urgent - Medical Device Correction letter dated August 1, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taked. The letter contained important information for the continued safe and proper use of your equipment. The letter stated that, a Field Service Engineer will visit each affected site to install the hardware upgrade. For further information or support concerning this issue contact your local Philips representative or local Philips Health office. For North America and Canada contact Customer Care Solution Center (1-800-722-9377), option 5: Enter Site ID or Follow the prompts).
Quantity
126 units
Distribution
Worldwide Distribution - US Nationwide including the states of: AL, AR, AZ, CA, DC, FL, GA, KY, MD, MI, MN, MS, ND, NM, NY, OH, OK, SC, TN, TX, VT, WA, and WI and the countries of: Austria, China, Denmark, France, Germany, Iceland, Japan, Netherlands, South Africa, Sweden, Switzerland, United Kingdom. There were two government consignees. There were no Canadian consignees.
Lot, serial or model codes
System Code # 728243, Serial #s: 7392, 7394, 7399, 7405, 7406, 7429, 7433, 7440, 7457, 7465, 7469, 7479, 7481, 7487, 7488, 7491, 7502, 7506, 7507, 7517, 7518, 7520, 7521, 7532, 7539, 7545, 7547, and 7556.  System Code # 728244, Serial #s: 7190, 7195, 7318, 7383, 7427, 7430, 7510, and 7533.  System Code #728256, Serial #s: 3163, 3500, 3500, 5623, 5655, 5655, 5666, 5670, 5715, 5724, 5747, 5773, 5776, 5813, 5836, 5855, 5862, 5868, 5877, 5897, 5911, 5911, 5916, 5935, 5969, and 5997.  System Code #728251, Serial #: 4051.  System Code # 728246, Serial #s: 5547, 5585, 5791, 5803, 5876, 5974.  System Code # 728235, Serial #s: 9233, 9234, 9261, 95022.  System Code # 728231, Serial #s: 9867, 9925, 9931, 9980, 9996, 10009, 10011, 10015, 10028, 10036, 10042, 10045, 10050, 10052, 10062, 10063, 10064, 10077, 10079, 10081, 10096, 10099, 10102, 10111, 10114, 10120, 10122, 10133, 10135, 10136, 10137, 10139, 10145, 10148, 10149, 10151, 10153, 10166, 10170, 10172, 10198, 10203, 95109, 95348, 95563.
510(k) numbers
["K012009","K033326","K033357"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register