Skip to the content

FDA Medical Device Recalls (openFDA)

Cardinal Health 200, LLC: Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201

Recall number
Z-2244-2021
Recalling firm
Cardinal Health 200, LLC
Product
Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201
Status
Terminated
Initiated
2021-06-16
Posted
2021-08-13
Terminated
2023-01-04
Reason
The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.
Root cause
Labeling Change Control
Action
Cardinal Health issued Urgent Medical Device Correction letter and Acknowledgement Form on June 16, 2021 to the direct consignees. Letter states reason for recall, health risk and action to take: Actions Required: 1) REVIEW the IFU for the Safety Scalpel N11 (vendor part number 73-1811). 2) COMMUNICATE the IFU to all personnel that utilize the Argyle" UVC Insertion Tray (SKU 43201) containing Safety Scalpel N11 (vendor part number 73-1811) 3) NOTIFY any customers to whom you may have distributed/forwarded affected product or will send the product about this product correction notice and share a copy of this notice and the attached IFU. 4) POST a copy of this notification in the location where the product is stored. 5) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to GMB-FieldCorrectiveAction@cardinalhealth.com whether or not you have affected product. Available Assistance: Please contact the Customer Service group for any questions or to arrange for credit and return of any product: " Hospital 800-964-5227 " Federal Government 800-444-1166 " Distributor 800-635-6021 " All other customers 888-444-5440 For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Quantity
7197 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
All lot codes
510(k) numbers
not on file
PMA numbers
not on file
Product code
FOS
Firm FEI
3014524790
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register