Skip to the content

FDA Medical Device Recalls (openFDA)

Bayer Healthcare: Angiographic Injector and Syringe, MEDRAD MRXperion MR Injection System, MEDRAD MRXperion Sterile Disposable MRI Kit Usage: The MEDRAD MRXperion MR Injection System is a syringe based fluid delivery system indicated for delivery of contrast media and saline during MR applications. It is intended to be used for the specific purpose of injecting intravenous MR contrast media and saline into the human vascular system for diagnostic studies in magnetic resonance imaging (MRI) applications with MRI scanners that have a magnetic field strength between 0.7 and 3.0 Tesla. Only trained healthcare professionals are intended to operate this device.

Recall number
Z-2244-2016
Recalling firm
Bayer Healthcare
Product
Angiographic Injector and Syringe, MEDRAD MRXperion MR Injection System, MEDRAD MRXperion Sterile Disposable MRI Kit Usage: The MEDRAD MRXperion MR Injection System is a syringe based fluid delivery system indicated for delivery of contrast media and saline during MR applications. It is intended to be used for the specific purpose of injecting intravenous MR contrast media and saline into the human vascular system for diagnostic studies in magnetic resonance imaging (MRI) applications with MRI scanners that have a magnetic field strength between 0.7 and 3.0 Tesla. Only trained healthcare professionals are intended to operate this device.
Status
Terminated
Initiated
2016-06-09
Posted
not on file
Terminated
2016-12-05
Reason
Bayer Healthcare is initiating this recall due to complaints that were received from customer sites describing a 4205 error message when the injector is used with a 3T scanner.
Root cause
Device Design
Action
Bayer sent an Urgent Medical Device Field Corrective Action letter dated June 8, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to take the following steps should a 4205 error message is encountered. For questions contact Bayer Customer Support Team at 1-800-633-7237.
Quantity
42 units
Distribution
Worldwide and the US state of NEW YORK
Lot, serial or model codes
Catalog numbers: MRXP 200 Serial numbers: 100009, 100011, 100019, 100020, 100022, 100026, 100027, 100028, 100036, 100047, 100065, 100066, 100069, 100074, 100075, 100085, 100090, 100091, 100096, 100097, 100098, 100099, 100100, 100101, 100103, 100104, 100106, 100107, 100108, 100109, 100110, 100111, 100112, 100113, 100114, 100118, 100119, 100129, 100152, 100155, 100161, 100166
510(k) numbers
["K143538"]
PMA numbers
not on file
Product code
DXT
Firm FEI
2520313
Firm city
Indianola
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register