FDA Medical Device Recalls (openFDA)
bioMerieux, Inc.: BacT/ALERT - FA Reagent Bottle The BacT/ALERT Microbial Detection System is used to determine if microorganisms are present in blood or other normally sterile body fluid samples taken from a patient suspected of having bacteremial/fungemia. The BacT/ALERT System and culture bottles provide both a microbial detection system and a culture media with suitable nutritional and environmental conditions for organisms commonly encountered in blood infections and other normally sterile body fluid infections. An inoculated bottle is placed into the instrument where it is incubated and continuously monitored for the presence of microorganisms that will grow in the BacT/ALERT FA bottle. BacT/ALERT FA Culture Bottles are used with the BacT/ALERT Microbial Detection System in qualitative procedures for enhanced recovery and detection of aerobic and facultative anaerobic microorganisms (bacterial and fungi) from blood and other normally sterile body fluids.
- Recall number
- Z-2244-2011
- Recalling firm
- bioMerieux, Inc.
- Product
- BacT/ALERT - FA Reagent Bottle The BacT/ALERT Microbial Detection System is used to determine if microorganisms are present in blood or other normally sterile body fluid samples taken from a patient suspected of having bacteremial/fungemia. The BacT/ALERT System and culture bottles provide both a microbial detection system and a culture media with suitable nutritional and environmental conditions for organisms commonly encountered in blood infections and other normally sterile body fluid infections. An inoculated bottle is placed into the instrument where it is incubated and continuously monitored for the presence of microorganisms that will grow in the BacT/ALERT FA bottle. BacT/ALERT FA Culture Bottles are used with the BacT/ALERT Microbial Detection System in qualitative procedures for enhanced recovery and detection of aerobic and facultative anaerobic microorganisms (bacterial and fungi) from blood and other normally sterile body fluids.
- Status
- Terminated
- Initiated
- 2011-04-13
- Posted
- 2011-05-24
- Terminated
- 2012-04-03
- Reason
- The charcoal suspension in a portion of the lot does not meet the intended density.
- Root cause
- Other
- Action
- BioMerieux sent an Urgent Product Removal Notice dated April 13, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Evaluate positive results adhering to their institution's protocols for determining false versus true positives. Negative bottles should be checked by smear and/or subculture at some point prior to discarding as negative. Any remaining inventory of BacT/ALERT FA culture bottle lot 1027481 should be destroyed in accordance with their facility's policies and procedures. An Acknowledgement and Product Replacement Form was included to be completed and returned to bioMerieux or credit or replacement of unused bottles.
- Quantity
- 327,700 bottles
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of China, Tokyo, South Korea, Canada, Sweden, Austria, Poland, and Belgium
- Lot, serial or model codes
- Catalog number: 259791, Lot umber 1027481, Expiry: December 8, 2011
- 510(k) numbers
- ["K020813"]
- PMA numbers
- not on file
- Product code
- MDB
- Firm FEI
- 3002769706
- Firm city
- Durham
- Firm state
- NC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28