Med Tec Inc: Type S Extension For use with Varian, CIVCO, Prescription Only - Product Usage: Device is intended for patient positioning.

Recall number
Z-2243-2020
Recalling firm
Med Tec Inc
Product
Type S Extension For use with Varian, CIVCO, Prescription Only - Product Usage: Device is intended for patient positioning.
Status
Terminated
Initiated
2020-05-12
Posted
2020-06-02
Terminated
2021-03-25
Reason
Potential for disengagement of the Varian Style Type-S Extension during patient repositioning. The risk associated with the disengagement of the extension could be a patient fall if the patient is not supported by clinical staff.
Root cause
Component design/selection
Action
On May 14, 2020, the firm sent a notification letter titled, "Urgent: Field Safety Notice Customer/Distributor Communication," to customers and distributors. The customer notification stated that a new warning will be added to the instruction manual at the next revision. The warning will state: To prevent extension from unintended disengagement, do not lift patient vertically to reposition while immobilized under a thermoplastic mask. Extensions that were further distributed, should be notified. An acknowledgement form should be returned to the firm by 05/29/2020. In addition, customers need to complete a form which states that they have been informed of the field safety notice involving the Type-S Extension, Varian Style and that notice has been read and understood by all operators of the device. In addition, the customers need to review their inventory for the affected products. Customers need to identify whether they no longer have the Type-S Extension, Varian Style or currently have the Type-S Extension, Varian Style. Customers need to acknowledge that have read and understand the Field Safety Notification and notified users of the device to not lift the patient vertically to reposition while immobilized under a thermoplastic mask. Customers need to identify the quantity of the devices at the facility, as well as, lot numbers and serial numbers. The forms should be returned by FAX or email to: Attention: CIVCO Radiotherapy Type-S Extension North American Customer use: Order@CivcoRT.com or 1-800-842-8688 International Customers use: DistorderCivcoRT.com or 1-712-737-8688 Customers can contact the firm's sales representative for additional assistance by calling the firm between 8:00AM and 5:00PM CST. The firm's distributor notification letters stated that the firm has initiated a field action due to the potential for disengagement of the Type-S Extension, Varian Style during patient setup and/or repositioning. The firm has iden
Quantity
171 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of WI, TX, AZ, FL, NE, TN, MA, FL, GA, IN, OR, PA, IL, LA, NH, MS, MI, OH, MN, VA, IA, CA, NC, SC, AK, UT, WA, MO, MT, KS, and WV. The countries of Africa, Colombia, Saudi Arabia, Costa Rica, South Korea, Hungary, Spain, and Italy.
Lot, serial or model codes
Part Number: MTCFHNV UDI: 00841439104156 Lot Numbers: M027990; M505880; M847500; M166160; M505890; M858580; M198170; M505900; M858590; M241650; M505910; M869030; M249180; M511240; M885500; M257630; M530920; M918000; M278760; M541350; M919980; M282330; M553300; M919990; M283380; M553720; M928400; M293670; M553740; M928410; M300750; M596220; M934260; M303010; M596230; M936110; M314410; M616690; M942400; M321880; M616700; M942510; M323560; M649630; M944110; M326580; M649670; M951030; M333420; M649770; M964730; M350700; M715520; M979830; M356130; M715540; M987460; M362290; M743010; M995860; M365740; M767820; M369700; M774760; M371990; M782390; M418080; M796350; M436620; M796360; M439700; M802120; M442410; M805650; M446460; M809050; M448260; M809060; M457870; M820300; M457880; M829290; M462640; M834400; M493710; M844990; M495030; M845000
510(k) numbers
["K933227"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1000143606
Firm city
Orange City
Firm state
IA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register