Biomet, Inc.: Vanguard 360 Femoral Boss Reamer, Label: REF 32-488474 VANGUARD 360 1NSTRUMENT FEMORAL BOSS REAMER NON-STERILE, BIOMET ORTHOPEDICS 56 EAST BELL DRIVE P.O. BOX 587 WARSAW, IN 46581 USA

Recall number
Z-2242-2014
Recalling firm
Biomet, Inc.
Product
Vanguard 360 Femoral Boss Reamer, Label: REF 32-488474 VANGUARD 360 1NSTRUMENT FEMORAL BOSS REAMER NON-STERILE, BIOMET ORTHOPEDICS 56 EAST BELL DRIVE P.O. BOX 587 WARSAW, IN 46581 USA
Status
Terminated
Initiated
2014-07-01
Posted
2014-08-15
Terminated
2015-10-30
Reason
The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.
Root cause
Process change control
Action
Urgent Medical Device Recall Notices dated July 15, 2014 were mailed with verification of delivery via FedEx. Response forms were provided to all consignees in order for Biomet to track returns. The letters informed the consignees to return the devices to Biomet and conduct a sub-recall sending the recall letter to all customers.
Quantity
461
Distribution
California, IA, FL, MD, UT, SC, AL, IL, CO, MA, KS, GA, WA, NV, AR, NH, OK,Pennsylvania, New York, Texas, Virginia, New Jersey, North Carolina, Kentucky, Michigan, Ohio, Wisconsin, Louisiana, Arizona, Indiana, South Dakota, Missouri & Montana. In addition to Korea, B.V., Japan, Puerto Rico, UK, Chile, New Zealand, Australia, & Canada
Lot, serial or model codes
Catalog Number: 32-48847 4 Lot Numbers: 022120, 022180, 162530, 232460, 475500 and 475510. 699240, 793350, 793570 Additional Scope: 065030, 095780, 164290, 235870, 253670, 273630, 308620, 418970, 420030, 421210, 570460, 585260, 619690, 708240, 711430, 726490, 816160, 896810, 929690, 953540, 968950
510(k) numbers
["K093293"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register