Skip to the content

FDA Medical Device Recalls (openFDA)

Dental Equipment LLC: The magnetic headrest is designed to fit the following dental chairs (THE HEADREST IS THE RECALLED ITEM, NOT THE CHAIRS): The dental chairs are sold under the brand names Marus and DCI Equipment. The affected models are: Marus MaxStar-Agency Model # DC1490; Marus MaxStar-Agency Model # DC1690; Marus NuStar-Agency Model # DC1700; Marus NuStar-Agency Model # DC1702; Marus ProStar-Agency Model # DC1540; Marus ProStar-Agency Model # DC1535; DCI Equipment DC1235 Hydraulic-Agency Model # DC1235; and DCI Equipment DC1335 Electromechanical-Agency Model # DC1335.

Recall number
Z-2241-2010
Recalling firm
Dental Equipment LLC
Product
The magnetic headrest is designed to fit the following dental chairs (THE HEADREST IS THE RECALLED ITEM, NOT THE CHAIRS): The dental chairs are sold under the brand names Marus and DCI Equipment. The affected models are: Marus MaxStar-Agency Model # DC1490; Marus MaxStar-Agency Model # DC1690; Marus NuStar-Agency Model # DC1700; Marus NuStar-Agency Model # DC1702; Marus ProStar-Agency Model # DC1540; Marus ProStar-Agency Model # DC1535; DCI Equipment DC1235 Hydraulic-Agency Model # DC1235; and DCI Equipment DC1335 Electromechanical-Agency Model # DC1335.
Status
Terminated
Initiated
2010-03-23
Posted
2010-08-20
Terminated
2012-01-04
Reason
The magnetic headrest on certain dental chairs may affect the function/programming of some implantable pacemakers or defibrillators if the implanted device is programmed to respond to a magnet.
Root cause
Other
Action
DCI Equipment sent an "Urgent Field Safety Alert" letter dated March 15, 2010 to consignees advising them of the potential problem. The letter instructed that patients with implantable pacemakers or defibrillators should avoid dental chairs with the magnetic headrest. A warning label was sent with the letter for insertion into the product Use and Care manual for operator warning. Consignees were asked to complete a Field Safety Alert Acknowledgement/Return Form and fax to the firm. Consignees were requested to notify customers.Consignees may contact the firm at 503 537-3602.
Quantity
108
Distribution
Worldwide distribution in the USA and Canada.
Lot, serial or model codes
The individual headrests do not have serial numbers. The headrest has a part number of "40R501". The part number is followed by a dash and a 2 or 3 digit number that indicates the color of the vinyl covering.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NRU
Firm FEI
1000523896
Firm city
Newberg
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register