Skip to the content

FDA Medical Device Recalls (openFDA)

Nobel Biocare USA LLC: NobelReplace Tapered Groovy RP 4.3x10mm. These products are root-form endosseous implants intended to be surgically placed in the bone of the upper or lower jaw arches to support for prosthetic devices.

Recall number
Z-2240-2008
Recalling firm
Nobel Biocare USA LLC
Product
NobelReplace Tapered Groovy RP 4.3x10mm. These products are root-form endosseous implants intended to be surgically placed in the bone of the upper or lower jaw arches to support for prosthetic devices.
Status
Terminated
Initiated
2007-11-02
Posted
2008-09-19
Terminated
2008-09-22
Reason
This recall was initiated after it was discovered that the NobelReplace Tapered Groovy RP 4.3x10mm, REF 32216 Lot#403504 has an incorrect cap label. The affected lot has a cap label identifying the implant as "4.3x8". Since this cap label may be the only means used for identifying the implant prior to surgery, it is possible that a user would try to use the 10mm implant as an 8mm implant.
Root cause
Labeling mix-ups
Action
The recall was initiated on 11/02/2007 with telephone calls to the affected customers. A "Urgent Medical Device Recall" letter with attached Customer Response form was then sent via US Postal Service beginning on 11/02/2007. The letter instructed customers to please immediately remove and return REF 32216 Lot #403504, NobelReplace Tapered Groovy RP 4.3x10mm. The firm contracted UPS courier service to pick up the recalled products. Customers were also instructed if the have further distributed the affected product to please immediately notify those customer to cease using this particular lot and forward the customer this notice. The firm indicated in the recall notice that they are in the process of sending out replacement products. Contact Kam Leung at 714-282-4800 if you have questions.
Quantity
197
Distribution
AR, AZ, CA, CO, FL, GA, IA, IL, MA, ME, MI, MO, NC, NC, NJ, NY, OK, PA, TX, UT, VA, WA
Lot, serial or model codes
Ref 32216 lot number: 403504
510(k) numbers
["K050258","K050705"]
PMA numbers
not on file
Product code
DZE
Firm FEI
1000125598
Firm city
Yorba Linda
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register