MICROVENTION INC.: FRED Flow Re-Direction Endoluminal Device, REF: FRED3507-PMA, FRED3511, FRED3511-PMA, FRED3516-CA, FRED3516-PMA, FRED3524, FRED3524-CA, FRED3524-PMA, FRED3536-CA, FRED3536-PMA, FRED4007-PMA, FRED4012-CA, FRED4012-PMA, FRED4017-PMA, FRED4026, FRED4026-PMA, FRED4038-PMA, FRED4508, FRED4508-PMA, FRED4513-PMA, FRED4518-PMA, FRED4528, FRED4528-PMA, FRED4539, FRED4539-CA, FRED4539-PMA, FRED5009, FRED5009-PMA, FRED5014-PMA, FRED5019-PMA, FRED5029, FRED5029-CA, FRED5029-PMA, FRED5514-CA, FRED5514-PMA, FRED5526-CA, FRED5526-PMA, MV-F351127, MV-F352427, MV-F401227, MV-F401727, MV-F451327, MV-F500927, MV-F501427, MV-F501927, MV-F552627

Recall number
Z-2238-2023
Recalling firm
MICROVENTION INC.
Product
FRED Flow Re-Direction Endoluminal Device, REF: FRED3507-PMA, FRED3511, FRED3511-PMA, FRED3516-CA, FRED3516-PMA, FRED3524, FRED3524-CA, FRED3524-PMA, FRED3536-CA, FRED3536-PMA, FRED4007-PMA, FRED4012-CA, FRED4012-PMA, FRED4017-PMA, FRED4026, FRED4026-PMA, FRED4038-PMA, FRED4508, FRED4508-PMA, FRED4513-PMA, FRED4518-PMA, FRED4528, FRED4528-PMA, FRED4539, FRED4539-CA, FRED4539-PMA, FRED5009, FRED5009-PMA, FRED5014-PMA, FRED5019-PMA, FRED5029, FRED5029-CA, FRED5029-PMA, FRED5514-CA, FRED5514-PMA, FRED5526-CA, FRED5526-PMA, MV-F351127, MV-F352427, MV-F401227, MV-F401727, MV-F451327, MV-F500927, MV-F501427, MV-F501927, MV-F552627
Status
Open, Classified
Initiated
2023-06-07
Posted
2023-07-27
Terminated
not on file
Reason
Intracranial aneurysm flow diverter device may be manufactured with an incorrect inner stent length and may not perform as intended.
Root cause
Process control
Action
On 6/7/23 recall notices were emailed to customers who were asked to stop use, quarantine, and return affected devices. Follow-up notices were mailed to customers on 6/16/23. Customers were asked to share the notice with all device users within the facility and network and share the notice with any organization where potentially affected devices have been transferred. In addition, customers were asked to complete and return the customer acknowledgement form. Inquiries can be directed to customer service at 800-990-8368.
Quantity
474
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, NY, FL, CA, PA, SC, TX, IL, MA, NJ, VA, GA, OH, IN, MI, KY, AL, MO, SD, LA, WA, HI, KS, MD, OK, DE, NC, MS and the countries of Taiwan, Chile, India, South Korea, Malaysia, Canada, France, Morocco, Belgium, Spain, Turkey, Poland, Russian Federation, Czech Republic, Slovenia, Germany, United Kingdom, El Salvador, United Arab Emirates, Bulgaria, Portugal, Brazil, Iran, Malta, Vietnam, South Africa, Jordan, Japan, Uruguay.
Lot, serial or model codes
Devices manufactured between June 1, 2020, and September 29, 2020. REF/UDI/Lot: FRED3507-PMA/00842429106655/200703556, 20070654P, 20072255L, 20072255N, 20073056A, 20073056B, 200807599, 20081453X, 20082656F, 20082656G, 20082852H, 20083152E; FRED3511/20073054C, 20091654Z, 200917511; FRED3511-PMA/00842429106662/20070754K, 200716562, 200721573, 20072252J, 20072255P, 20072255Q, 20080851B, 20082656H, 20082656J, 20090152R; FRED3516-CA/20081154U; FRED3516-PMA/00842429106679/20070654Q, 200716563, 20072158F, 200731555, 20080354B, 20081154W; FRED3524/00811425020982/200702553, 200706563, 20071557U, 200720563, 20081955R, 20091654T, 20091857N, 20092357X; FRED3524-CA/200924555; FRED3524-PMA/00842429106686/200716564, 200716565, 20072255U, 20072255V, 2007225AZ, 20073056C, 20073056Q, 20073056R, 20080459V, 20080558E, 20080853W, 2008115AC, 20081255W, 20081354Q, 200909564, 200909565, 200909664, 20091558Z, 20092156J, 20092166J; FRED3536-CA/20092859F, 20092955K; FRED3536-PMA/20070654R, 20071556Z, 200716566, 20072255W, 20072255X, 20080558G, 20081255X, 20090955W, 200915547, 200915591, 200916569, 200923539, 20092356W, 20092356X, 20092957Y; FRED4007-PMA/00842429106709/20072255Y, 20072255Z, 20072455H, 20073056Y, 20073056Z, 200730571, 20080558H, 20080558K, 20090255J, 20091555N; FRED4012-CA/20081255H; FRED4012-PMA/00842429106723/200703557, 200703575, 200730572, 200730573, 200806569, 20081957L, 20082656Q, 20092356Y, 20092356Z, 200929581; FRED4017-PMA/00842429106730/20070154P, 200702514, 200703558, 200730575, 200806589, 200812562, 20081354G, 200813554, 200901548, 200902548, 2009085AT, 2009095B2; FRED4026/20070754G, 20080556Z, 20081356V, 20081955T, 200928541M; FRED4026-PMA/00842429106747/20070452Q, 200730577, 20081453Y, 200923571, 200923572, 200927516, 200928532, 200928541, 20093055H, 20093065H; FRED4038-PMA/00842429106754/200722561, 200722562, 20082656R, 20083152F, 200831594, 200901595, 200902545, 200902546, 20090759V, 200909569, 20091656E, 20091659G, 2009175A1, 2009175A3, 20091953C, 20091953E, 20091953F, 20092156N, 20092156R, 2009225AX, 200923573, 200923574, 20092459J, 20092469J, 20092557J, 200926535, 20092956H, 20093056X, 200930586; FRED4508/00811425021057/200805571, 20080851A, 20081154R; FRED4508-PMA/00842429106761/200703559, 200722563, 200722564, 20073057C, 20073057E, 20082656V, 20082656W, 200831598, 2009015A1, 20091555Q; FRED4513-PMA/00842429106778/200702569, 20070256A, 200713549, 20073057F, 20073057G, 200807546, 200902562, 200923575, 200923576; FRED4518-PMA/00842429106785/200702597, 20070355A, 200721576, 20073057J, 20080851E, 200808535; FRED4528/20070653Z, 200716556, 20072258F; FRED4528-PMA/200902564, 200908524; FRED4539/200706541, 20071553E, 20092955L; FRED4539-CA/20092853Z, 20092957K; FRED4539-PMA/00842429106808/200722565, 200722566, 20080558Q, 20080558R, 200812568, 200812569, 20081957V, 20081957W, 20082656X, 20082656Y, 20082656Z, 20082954E, 20083152G, 20083154W, 200901594, 200902565, 200902566, 2009025AQ, 2009025CB, 200903594, 20090451X, 20090956C, 20091055L, 20091253U, 20091656F, 20091659F, 20091751N, 20091952B, 20091953A, 200922537, 200922584, 200922585, 200922684, 200923577, 200923578, 200923678, 20092553C, 20092553E, 200925571, 20092557H, 20092563C, 200926522, 20092652C, 20092858U, 20092868U, 200929583; FRED5009/20081254Z, 20081355F; FRED5009-PMA/00842429106815/20070256B, 200703574, 200722567, 20073057P, 20073057Q, 20073057R, 20080457X, 20080558T, 20080558U, 20081258J, 200902567; FRED5014-PMA/00842429106822/20070256E, 20070256F, 200706571, 20071553L, 20071651L, 200717545, 2007225B1, 200811596, 20082559G; FRED5019-PMA/00842429106839/20070754L, 200721577, 20073057T, 200826571, 20090152T, 200901593, 2009015A6, 2009025BZ; FRED5029/20073054E, 200812551; FRED5029-CA/20091655V, 20091665V, 20092156Q, 20092166Q; FRED5029-PMA/00842429106846/20070256J, 20070256K, 20070256L, 20070452X, 200713556, 2007135A3, 20081957Y, 200922539, 200923579, 20092357A, 200928542, 200929584; FRED5514-CA/20071655K, 2009085AU, 200922536; FRED5514-PMA/00842429106853/20070256N, 20070452Y, 20072258C, 20081957Z, 200819581, 20082655Z, 200826561, 200826661, 20082758U, 20090154H, 200902568, 200904552, 20090457F, 20090956G; FRED5526-CA/20092957L; FRED5526-PMA/00842429106860/20070254L, 20070256P, 20070256Q, 20070256R, 20071557R, 200819583, 200826573, 20092156L, 20092357B, 200929585; MV-F351127/20081358X, 20090954X, 20090956K; MV-F352427/200722M5U, 200730M6Q; MV-F401227/200730M72, 200806M69, 200819M7L, 200826M6Q, 200923M6Z; MV-F401727/200701M4P, 200901M48, 200909MB2; MV-F451327/200702M69, 200702M6A, 200713M49, 200730M7F, 200730M7G, 200807M46, 200923M75; MV-F500927/20090956N; MV-F501427/200702M6F, 200706M71, 200715M3L, 200716M1L, 200811M96; MV-F501927/200730M7T; MV-F552627/200702M4L, 200702M6Q
510(k) numbers
not on file
PMA numbers
not on file
Product code
OUT
Firm FEI
3013556777
Firm city
Aliso Viejo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register