MICROVENTION INC.: FRED Flow Re-Direction Endoluminal Device, REF: FRED3507-PMA, FRED3511, FRED3511-PMA, FRED3516-CA, FRED3516-PMA, FRED3524, FRED3524-CA, FRED3524-PMA, FRED3536-CA, FRED3536-PMA, FRED4007-PMA, FRED4012-CA, FRED4012-PMA, FRED4017-PMA, FRED4026, FRED4026-PMA, FRED4038-PMA, FRED4508, FRED4508-PMA, FRED4513-PMA, FRED4518-PMA, FRED4528, FRED4528-PMA, FRED4539, FRED4539-CA, FRED4539-PMA, FRED5009, FRED5009-PMA, FRED5014-PMA, FRED5019-PMA, FRED5029, FRED5029-CA, FRED5029-PMA, FRED5514-CA, FRED5514-PMA, FRED5526-CA, FRED5526-PMA, MV-F351127, MV-F352427, MV-F401227, MV-F401727, MV-F451327, MV-F500927, MV-F501427, MV-F501927, MV-F552627
- Recall number
- Z-2238-2023
- Recalling firm
- MICROVENTION INC.
- Product
- FRED Flow Re-Direction Endoluminal Device, REF: FRED3507-PMA, FRED3511, FRED3511-PMA, FRED3516-CA, FRED3516-PMA, FRED3524, FRED3524-CA, FRED3524-PMA, FRED3536-CA, FRED3536-PMA, FRED4007-PMA, FRED4012-CA, FRED4012-PMA, FRED4017-PMA, FRED4026, FRED4026-PMA, FRED4038-PMA, FRED4508, FRED4508-PMA, FRED4513-PMA, FRED4518-PMA, FRED4528, FRED4528-PMA, FRED4539, FRED4539-CA, FRED4539-PMA, FRED5009, FRED5009-PMA, FRED5014-PMA, FRED5019-PMA, FRED5029, FRED5029-CA, FRED5029-PMA, FRED5514-CA, FRED5514-PMA, FRED5526-CA, FRED5526-PMA, MV-F351127, MV-F352427, MV-F401227, MV-F401727, MV-F451327, MV-F500927, MV-F501427, MV-F501927, MV-F552627
- Status
- Open, Classified
- Initiated
- 2023-06-07
- Posted
- 2023-07-27
- Terminated
- not on file
- Reason
- Intracranial aneurysm flow diverter device may be manufactured with an incorrect inner stent length and may not perform as intended.
- Root cause
- Process control
- Action
- On 6/7/23 recall notices were emailed to customers who were asked to stop use, quarantine, and return affected devices. Follow-up notices were mailed to customers on 6/16/23. Customers were asked to share the notice with all device users within the facility and network and share the notice with any organization where potentially affected devices have been transferred. In addition, customers were asked to complete and return the customer acknowledgement form. Inquiries can be directed to customer service at 800-990-8368.
- Quantity
- 474
- Distribution
- Worldwide - US Nationwide distribution including in the states of AZ, NY, FL, CA, PA, SC, TX, IL, MA, NJ, VA, GA, OH, IN, MI, KY, AL, MO, SD, LA, WA, HI, KS, MD, OK, DE, NC, MS and the countries of Taiwan, Chile, India, South Korea, Malaysia, Canada, France, Morocco, Belgium, Spain, Turkey, Poland, Russian Federation, Czech Republic, Slovenia, Germany, United Kingdom, El Salvador, United Arab Emirates, Bulgaria, Portugal, Brazil, Iran, Malta, Vietnam, South Africa, Jordan, Japan, Uruguay.
- Lot, serial or model codes
- Devices manufactured between June 1, 2020, and September 29, 2020. REF/UDI/Lot: FRED3507-PMA/00842429106655/200703556, 20070654P, 20072255L, 20072255N, 20073056A, 20073056B, 200807599, 20081453X, 20082656F, 20082656G, 20082852H, 20083152E; FRED3511/20073054C, 20091654Z, 200917511; FRED3511-PMA/00842429106662/20070754K, 200716562, 200721573, 20072252J, 20072255P, 20072255Q, 20080851B, 20082656H, 20082656J, 20090152R; FRED3516-CA/20081154U; FRED3516-PMA/00842429106679/20070654Q, 200716563, 20072158F, 200731555, 20080354B, 20081154W; FRED3524/00811425020982/200702553, 200706563, 20071557U, 200720563, 20081955R, 20091654T, 20091857N, 20092357X; FRED3524-CA/200924555; FRED3524-PMA/00842429106686/200716564, 200716565, 20072255U, 20072255V, 2007225AZ, 20073056C, 20073056Q, 20073056R, 20080459V, 20080558E, 20080853W, 2008115AC, 20081255W, 20081354Q, 200909564, 200909565, 200909664, 20091558Z, 20092156J, 20092166J; FRED3536-CA/20092859F, 20092955K; FRED3536-PMA/20070654R, 20071556Z, 200716566, 20072255W, 20072255X, 20080558G, 20081255X, 20090955W, 200915547, 200915591, 200916569, 200923539, 20092356W, 20092356X, 20092957Y; FRED4007-PMA/00842429106709/20072255Y, 20072255Z, 20072455H, 20073056Y, 20073056Z, 200730571, 20080558H, 20080558K, 20090255J, 20091555N; FRED4012-CA/20081255H; FRED4012-PMA/00842429106723/200703557, 200703575, 200730572, 200730573, 200806569, 20081957L, 20082656Q, 20092356Y, 20092356Z, 200929581; FRED4017-PMA/00842429106730/20070154P, 200702514, 200703558, 200730575, 200806589, 200812562, 20081354G, 200813554, 200901548, 200902548, 2009085AT, 2009095B2; FRED4026/20070754G, 20080556Z, 20081356V, 20081955T, 200928541M; FRED4026-PMA/00842429106747/20070452Q, 200730577, 20081453Y, 200923571, 200923572, 200927516, 200928532, 200928541, 20093055H, 20093065H; FRED4038-PMA/00842429106754/200722561, 200722562, 20082656R, 20083152F, 200831594, 200901595, 200902545, 200902546, 20090759V, 200909569, 20091656E, 20091659G, 2009175A1, 2009175A3, 20091953C, 20091953E, 20091953F, 20092156N, 20092156R, 2009225AX, 200923573, 200923574, 20092459J, 20092469J, 20092557J, 200926535, 20092956H, 20093056X, 200930586; FRED4508/00811425021057/200805571, 20080851A, 20081154R; FRED4508-PMA/00842429106761/200703559, 200722563, 200722564, 20073057C, 20073057E, 20082656V, 20082656W, 200831598, 2009015A1, 20091555Q; FRED4513-PMA/00842429106778/200702569, 20070256A, 200713549, 20073057F, 20073057G, 200807546, 200902562, 200923575, 200923576; FRED4518-PMA/00842429106785/200702597, 20070355A, 200721576, 20073057J, 20080851E, 200808535; FRED4528/20070653Z, 200716556, 20072258F; FRED4528-PMA/200902564, 200908524; FRED4539/200706541, 20071553E, 20092955L; FRED4539-CA/20092853Z, 20092957K; FRED4539-PMA/00842429106808/200722565, 200722566, 20080558Q, 20080558R, 200812568, 200812569, 20081957V, 20081957W, 20082656X, 20082656Y, 20082656Z, 20082954E, 20083152G, 20083154W, 200901594, 200902565, 200902566, 2009025AQ, 2009025CB, 200903594, 20090451X, 20090956C, 20091055L, 20091253U, 20091656F, 20091659F, 20091751N, 20091952B, 20091953A, 200922537, 200922584, 200922585, 200922684, 200923577, 200923578, 200923678, 20092553C, 20092553E, 200925571, 20092557H, 20092563C, 200926522, 20092652C, 20092858U, 20092868U, 200929583; FRED5009/20081254Z, 20081355F; FRED5009-PMA/00842429106815/20070256B, 200703574, 200722567, 20073057P, 20073057Q, 20073057R, 20080457X, 20080558T, 20080558U, 20081258J, 200902567; FRED5014-PMA/00842429106822/20070256E, 20070256F, 200706571, 20071553L, 20071651L, 200717545, 2007225B1, 200811596, 20082559G; FRED5019-PMA/00842429106839/20070754L, 200721577, 20073057T, 200826571, 20090152T, 200901593, 2009015A6, 2009025BZ; FRED5029/20073054E, 200812551; FRED5029-CA/20091655V, 20091665V, 20092156Q, 20092166Q; FRED5029-PMA/00842429106846/20070256J, 20070256K, 20070256L, 20070452X, 200713556, 2007135A3, 20081957Y, 200922539, 200923579, 20092357A, 200928542, 200929584; FRED5514-CA/20071655K, 2009085AU, 200922536; FRED5514-PMA/00842429106853/20070256N, 20070452Y, 20072258C, 20081957Z, 200819581, 20082655Z, 200826561, 200826661, 20082758U, 20090154H, 200902568, 200904552, 20090457F, 20090956G; FRED5526-CA/20092957L; FRED5526-PMA/00842429106860/20070254L, 20070256P, 20070256Q, 20070256R, 20071557R, 200819583, 200826573, 20092156L, 20092357B, 200929585; MV-F351127/20081358X, 20090954X, 20090956K; MV-F352427/200722M5U, 200730M6Q; MV-F401227/200730M72, 200806M69, 200819M7L, 200826M6Q, 200923M6Z; MV-F401727/200701M4P, 200901M48, 200909MB2; MV-F451327/200702M69, 200702M6A, 200713M49, 200730M7F, 200730M7G, 200807M46, 200923M75; MV-F500927/20090956N; MV-F501427/200702M6F, 200706M71, 200715M3L, 200716M1L, 200811M96; MV-F501927/200730M7T; MV-F552627/200702M4L, 200702M6Q
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OUT
- Firm FEI
- 3013556777
- Firm city
- Aliso Viejo
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28