Skip to the content

FDA Medical Device Recalls (openFDA)

Spectrum Medical Ltd.: Quantum Pump Console

Recall number
Z-2238-2019
Recalling firm
Spectrum Medical Ltd.
Product
Quantum Pump Console
Status
Terminated
Initiated
2019-05-29
Posted
2019-08-23
Terminated
2020-08-21
Reason
Potential for Quantum Pump Console, part of the Quantum Perfusion Systems, to unexpectedly shut down while in use.
Root cause
Software design
Action
Spectrum Medical notified customers on about 05/29/2019 via "URGENT FIELD SAFETY CORRECTION" letter. Customers were informed that Spectrum Medical is in the process of installing the most recent software update v6.2.3.16 to the Quantum Perfusion Systems, (Quantum Pump Console). This software update corrects an error that could potentially lead to power loss. All Quantum Perfusion Systems are in scope of this issue. Customers were instructed to be aware that user intervention may be necessary with continued use before the software update, assure that all users receive notice of this issue, confirm receipt of the Urgent Medical Device Correction by returning the customer response form, and a Spectrum Medical service team will contact each customer and arrange a time and date to update the systems with the most recent software update. Additionally, customers were instructed to notify all relevant personnel and to customers if the devices have been further distributed.
Quantity
108 units (8 US, 100 OUS)
Distribution
US distribution to NY, OH, VA, WA. International distribution to the Netherlands.
Lot, serial or model codes
Software version 6.2.1.9; Serial Numbers PH1000005, PH1000020, PH1000027, PH1000030, PH1000077, PH1000081, PH1000082, PH1000086, PH1000087, PH1000093, PH1000100, PH1000101, PH1000102, PH1000103, PH1000109, PH1000110, PH1000112, PH1000113, PH1000114, PH1000115, PH1000120, PH1000127, PH1000129, PH1000131, PH1000132, PH1000134, PH1000135, PH1000137, PH1000139, PH1000140, PH1000141, PH1000144, PH1000145, PH1000150, PH1000156, PH1000158, PH1000159, PH1000160, PH1000161, PH1000162, PH1000165, PH1000168, PH1000170, PH1000171, PH1000172, PH1000175, PH1000176, PH1000177, PH1000178, PH1000179, PH1000180, PH1000182, PH1000201, PH1000208, PH1000209, PH1000210, PH1000211, PH1000212, PH1000213, PH1000215, PH1000233, PH1000256, PH1000257, PH1000259, PH1000438, PH1000439, PH1000440, PH1000441, PH1000442, PH1000445, PH1000455, PH1000495, PH1000498, PH1000522, PH1000615, PH1000644, PH1000645, PH1000646, PH1000647, PH1000648, PH1000649, PH1000668, PH1000669, PH1000670, PH1000682, PH1000683, PH1000684, PH1000686, PH1000687, PH1000688, PH1000689, PH1000692, PH1000722, PH1000726, PH1000727, PH1000740, PH1000742, PH1000743, PH1000744, PH1000879, PH1000880, PH1000882, PH1000885, PH1000886, PH1000895, PH1000897, PH1000939, PH1000940
510(k) numbers
["K173834"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
3006073153
Firm city
Gloucester
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.