FDA Medical Device Recalls (openFDA)
Spectrum Medical Ltd.: Quantum Pump Console
- Recall number
- Z-2238-2019
- Recalling firm
- Spectrum Medical Ltd.
- Product
- Quantum Pump Console
- Status
- Terminated
- Initiated
- 2019-05-29
- Posted
- 2019-08-23
- Terminated
- 2020-08-21
- Reason
- Potential for Quantum Pump Console, part of the Quantum Perfusion Systems, to unexpectedly shut down while in use.
- Root cause
- Software design
- Action
- Spectrum Medical notified customers on about 05/29/2019 via "URGENT FIELD SAFETY CORRECTION" letter. Customers were informed that Spectrum Medical is in the process of installing the most recent software update v6.2.3.16 to the Quantum Perfusion Systems, (Quantum Pump Console). This software update corrects an error that could potentially lead to power loss. All Quantum Perfusion Systems are in scope of this issue. Customers were instructed to be aware that user intervention may be necessary with continued use before the software update, assure that all users receive notice of this issue, confirm receipt of the Urgent Medical Device Correction by returning the customer response form, and a Spectrum Medical service team will contact each customer and arrange a time and date to update the systems with the most recent software update. Additionally, customers were instructed to notify all relevant personnel and to customers if the devices have been further distributed.
- Quantity
- 108 units (8 US, 100 OUS)
- Distribution
- US distribution to NY, OH, VA, WA. International distribution to the Netherlands.
- Lot, serial or model codes
- Software version 6.2.1.9; Serial Numbers PH1000005, PH1000020, PH1000027, PH1000030, PH1000077, PH1000081, PH1000082, PH1000086, PH1000087, PH1000093, PH1000100, PH1000101, PH1000102, PH1000103, PH1000109, PH1000110, PH1000112, PH1000113, PH1000114, PH1000115, PH1000120, PH1000127, PH1000129, PH1000131, PH1000132, PH1000134, PH1000135, PH1000137, PH1000139, PH1000140, PH1000141, PH1000144, PH1000145, PH1000150, PH1000156, PH1000158, PH1000159, PH1000160, PH1000161, PH1000162, PH1000165, PH1000168, PH1000170, PH1000171, PH1000172, PH1000175, PH1000176, PH1000177, PH1000178, PH1000179, PH1000180, PH1000182, PH1000201, PH1000208, PH1000209, PH1000210, PH1000211, PH1000212, PH1000213, PH1000215, PH1000233, PH1000256, PH1000257, PH1000259, PH1000438, PH1000439, PH1000440, PH1000441, PH1000442, PH1000445, PH1000455, PH1000495, PH1000498, PH1000522, PH1000615, PH1000644, PH1000645, PH1000646, PH1000647, PH1000648, PH1000649, PH1000668, PH1000669, PH1000670, PH1000682, PH1000683, PH1000684, PH1000686, PH1000687, PH1000688, PH1000689, PH1000692, PH1000722, PH1000726, PH1000727, PH1000740, PH1000742, PH1000743, PH1000744, PH1000879, PH1000880, PH1000882, PH1000885, PH1000886, PH1000895, PH1000897, PH1000939, PH1000940
- 510(k) numbers
- ["K173834"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 3006073153
- Firm city
- Gloucester
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28