Skip to the content

FDA Medical Device Recalls (openFDA)

Mckesson Medical Imaging Group: Horizon Medical Imaging system. Product Usage: Horizon Medical Imaging is a medical image and information management application that is intended to receive, transmit store, retrieve, display, print and process digital medical images, digital medical video, and associated medical information from various medical image systems. The medical modalities of these medical imaging systems include, but are not limited to, all modalities supported by ACR/NEMA DICOM 3.0 (specifically including mammographic images). Horizon Medical Imaging is intended to function with a variety of storage systems and printers via DICOM and other computer industry standard interfaces and protocols. Horizon Medical Imaging is indicated for use by trained medical professionals including, but not limited to, radiologist, physicians, and medical technologists. Horizon Medical Imaging is also indicated for use in soft copy diagnostic interpretation of medical images and video by physicians trained in such practice (specifically including soft copy diagnostic interpretation of mammographic images).

Recall number
Z-2238-2012
Recalling firm
Mckesson Medical Imaging Group
Product
Horizon Medical Imaging system. Product Usage: Horizon Medical Imaging is a medical image and information management application that is intended to receive, transmit store, retrieve, display, print and process digital medical images, digital medical video, and associated medical information from various medical image systems. The medical modalities of these medical imaging systems include, but are not limited to, all modalities supported by ACR/NEMA DICOM 3.0 (specifically including mammographic images). Horizon Medical Imaging is intended to function with a variety of storage systems and printers via DICOM and other computer industry standard interfaces and protocols. Horizon Medical Imaging is indicated for use by trained medical professionals including, but not limited to, radiologist, physicians, and medical technologists. Horizon Medical Imaging is also indicated for use in soft copy diagnostic interpretation of medical images and video by physicians trained in such practice (specifically including soft copy diagnostic interpretation of mammographic images).
Status
Terminated
Initiated
2012-05-29
Posted
2012-08-22
Terminated
2013-09-12
Reason
There is a potential to save reports in RadReport with an Incorrect Final Status.
Root cause
Software design
Action
McKesson sent an Urgent Field Safety Notice letter dated May 29, 2012 to all affected consignees. The letter identified the affected product, circumstances under which the issue occurs and immediate actions to be taken. Customers were advised to notify the Customer Support department that they have read and understood this Advisory Notice at 1-800-663-2533. For questions contact the Support Manager at the Customer Support department.
Quantity
516 units
Distribution
Worldwide Distribution - U.S. Nationwide and the countries of: Canada, France, Ireland and United Kingdom.
Lot, serial or model codes
Horizon Medical Imaging (HMI) versions 11.5, 11.5.1, 11.5.2, 11.6 and 11.7.
510(k) numbers
["K043146"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3003757840
Firm city
Canada
Firm state
British Columbia
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register