Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter, Inc.: Ringed DxI Reaction Vessels (RVs)

Recall number
Z-2237-2023
Recalling firm
Beckman Coulter, Inc.
Product
Ringed DxI Reaction Vessels (RVs)
Status
Open, Classified
Initiated
2023-06-20
Posted
2023-07-26
Terminated
not on file
Reason
Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
Root cause
Under Investigation by firm
Action
A customer letter titled, Urgent Medical Device Recall, dated 06/07/2023, was issued to the impacted consignee via first-class mail and email. The consignee is instructed to discard the product and request a replacement. The consignee is asked to complete the response form and submit to the recalling firm.
Quantity
213 cases of 10,000 RVs (2,130,000 RVs)
Distribution
Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI, GA, NC, FL, WV and the country of Qatar.
Lot, serial or model codes
UDI: 15099590229979; Catalog Number: 386167; Lot Numbers: L17855111, L17798598, L18046334
510(k) numbers
["K023764"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2122870
Firm city
Chaska
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register