FDA Medical Device Recalls (openFDA)
Zimmer Biomet, Inc.: Orthopedic Salvage System (OSS) Hybrid Poly Tibia Item # 151814
- Recall number
- Z-2235-2019
- Recalling firm
- Zimmer Biomet, Inc.
- Product
- Orthopedic Salvage System (OSS) Hybrid Poly Tibia Item # 151814
- Status
- Terminated
- Initiated
- 2019-07-02
- Posted
- not on file
- Terminated
- 2021-09-10
- Reason
- Cleaning processes potentially being ineffective
- Root cause
- Material/Component Contamination
- Action
- Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Certificate of Acknowledgement and send to Corporate Quality. PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed t
- Quantity
- 30
- Distribution
- Distributed to the following states in the USA: AZ, CA, DC, GA, IL, KY, LA, MA, MD, MN, NC, NJ, NM, OH,OK, OR, SC, TX. Distributed to Japan and Netherlands.
- Lot, serial or model codes
- Lot # All lots expiring prior to 06/30/2024 Affected products were distributed between June 2016 and February 2019.
- 510(k) numbers
- ["K142814"]
- PMA numbers
- not on file
- Product code
- JDI
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28