Skip to the content

FDA Medical Device Recalls (openFDA)

SEASPINE ORTHOPEDICS CORPORATION: Brand Name: Fathom Pedicle-Based Retractor Product Name: C/C Blade, 35-55mm Model/Catalog Number: MR2102035 Product Description: Cranial Caudal Blade Component: Yes, a component of a retractor blade

Recall number
Z-2234-2024
Recalling firm
SEASPINE ORTHOPEDICS CORPORATION
Product
Brand Name: Fathom Pedicle-Based Retractor Product Name: C/C Blade, 35-55mm Model/Catalog Number: MR2102035 Product Description: Cranial Caudal Blade Component: Yes, a component of a retractor blade
Status
Open, Classified
Initiated
2024-05-17
Posted
2024-06-28
Terminated
not on file
Reason
Cranial Caudal Blades manufactured incorrectly could result in fracture and/or damage to the bone structure
Root cause
Under Investigation by firm
Action
On May 17, 2024, SeaSpine Orthopedics Corporation issued a "Urgent: Medical Device Recall" notification. via E-Mail (DocuSign) to affected consignees. 1. Please return the entire Fathom set to SeaSpine s Carlsbad facility for evaluation and potential rework of the affected blades. A shipping label is attached for your convenience. 2. To support your upcoming cases, please continue to utilize the loaner pool. Upon receipt of the affected inventory, it will take approximately three weeks for rework and return to the loaner pool. 3. Once the inventory has been returned, please acknowledge via DocuSign.
Quantity
21 units
Distribution
United States Nationwide distribution in the states of Nevada, Texas, Michigan, Rhode Island, California, Oregon, Oklahoma, New York, Minnesota.
Lot, serial or model codes
UDI: 10889981300050 Lot Number: WT36298C
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAD
Firm FEI
3012120772
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register