FDA Medical Device Recalls (openFDA)
Oridion Medical 1987 Ltd.: MicroStream Filterline Infant/Neonatal products: Part # 006324 - FILTERLINE H SET INF/NEO 25 UN (Inf/Neo Airway Adapter). The adult/pediatric airway adapter and the adult/pediatric FilterLine and VitaLine sets are not affected by this recall. The intended use of Microstream Filter Lines is to convey a sample of a patient's expired breath to a capnograph in order to measure the CO2 partial pressure in intubated patients. It is a component of the FilterLine H Set Inf/Neo, Part # 006324 and the VitaLine H Set Inf/Neo, Part # 010807.
- Recall number
- Z-2234-2011
- Recalling firm
- Oridion Medical 1987 Ltd.
- Product
- MicroStream Filterline Infant/Neonatal products: Part # 006324 - FILTERLINE H SET INF/NEO 25 UN (Inf/Neo Airway Adapter). The adult/pediatric airway adapter and the adult/pediatric FilterLine and VitaLine sets are not affected by this recall. The intended use of Microstream Filter Lines is to convey a sample of a patient's expired breath to a capnograph in order to measure the CO2 partial pressure in intubated patients. It is a component of the FilterLine H Set Inf/Neo, Part # 006324 and the VitaLine H Set Inf/Neo, Part # 010807.
- Status
- Terminated
- Initiated
- 2011-03-25
- Posted
- 2011-05-25
- Terminated
- 2012-03-19
- Reason
- fine plastic strands on inner surface of the infant neonatal airway adapter may become dislodged and inhaled by the patient
- Root cause
- Equipment maintenance
- Action
- The firm, Oridion, sent an "Microstream Filterline Field Correction Notice" letter dated March 25, 2011 to its US and foreign customers via telephone, email with the documentation discussed in the call, and letter. The letter described the product, problem and actions to be taken. The customers were instructed to place all affected product held in their inventory on ship hold; gather these units in their inventory and at the customer sites and exchange with the new product; inspect filter line starter (sample) kits shipped with Capnostream 20 and Microcap patient monitors for affected product, if affected Infant/Neonatal filter lines are found, remove them from the starter (sample) kit; send a purchase order in order to obtain replacement product for units in their stock and at customers sites, and complete and return the Inventory Reporting Form Excel Spread Sheet within two weeks of receiving the email even if they or their customers do not have the affected inventory. Oridion shall conduct a removal of the products from the related production lots from the field and an immediate replacement with product from subsequent production lots that meet all requirements. This action will be conducted by Oridion's distributors or by Oridion directly in some cases. Oridion anticipates that new product will be available for shipment during the week of April 4th. Questions regarding the recall can be directed towards the Customer Service Coordinator at: 1-888-674-3466 x1234 or NA-FCA@oridion.com.
- Quantity
- 9,850 units in US (394 boxes with 25 units ea)
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- Lot Numbers: M8286M10 M8300M10 M8406P10 M8544B11 M8555B11 M8575B11
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- CCK
- Firm FEI
- 3001227743
- Firm city
- Jerusalem
- Firm state
- Har Hotzvim Industrial Park
- Firm country
- Israel
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28