Terumo Cardiovascular Systems Corporation: Interface module for CDl System 500 Product Usage: The Interface Module for CDl 500 is the interface between the CDl 500 monitor and the system. The module performs the following functions: Communicates primary pump flow data to the CDl 500 Receives blood parameter data from the CDl 500
- Recall number
- Z-2233-2012
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Interface module for CDl System 500 Product Usage: The Interface Module for CDl 500 is the interface between the CDl 500 monitor and the system. The module performs the following functions: Communicates primary pump flow data to the CDl 500 Receives blood parameter data from the CDl 500
- Status
- Terminated
- Initiated
- 2012-07-26
- Posted
- 2012-08-21
- Terminated
- 2014-06-27
- Reason
- Terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the odor of smoke was as associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the modu
- Root cause
- Device Design
- Action
- Terumo sent an URGENT MEDICAL DEVICE Recall letter dated July 24, 2012 to all affected customers. The letter identified the affected product, reason for correction, potential hazard and actions to be taken. The letter stated that Terumo CVS will replace all affected modules. Customers were informed that a Terumo VS Representative will contact the user to schedule a service call for the replacement upon receipt of the Response Form attached. For questions contact Terumo CVS Customer Service at 800-521-2818.
- Quantity
- 582 units
- Distribution
- Worldwide Distribution - US Nationwide including the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, and WV and the countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Germany, Hong Kong, Indonesia, Japan, Jordan, Malaysia, Singapore, South Africa, South Korea, Taiwan, Thailand, UNITED ARAB EMIRATES (UAE), and Vietnam
- Lot, serial or model codes
- catalog number: 803479 and serial numbers:00009-00122, 00124-00210, 00213-00365, 00367-00458, and 00461-00596.
- 510(k) numbers
- ["K022947"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28