FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: AXIOM Artis MP/Artis dMP and Artis zee Multipurpose. Angiographic x-ray system.
- Recall number
- Z-2233-2010
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- AXIOM Artis MP/Artis dMP and Artis zee Multipurpose. Angiographic x-ray system.
- Status
- Terminated
- Initiated
- 2010-07-13
- Posted
- 2010-08-16
- Terminated
- 2012-07-05
- Reason
- The Axiom Artis MP, Artis dMP or Artis zee Multipurpose C-arm gearbox can become damaged. If the gearbox is damaged, the C-arm may tilt away when being repositioned and collide with nearby objects in its range of motion.
- Root cause
- Other
- Action
- Siemens has issued an "Urgent Field Safety Notice: Customer Safety Information to consignees with Update Instructions AX033/10/S. This letter informed customers of the issue, and provides an Addendum to the User Manual. The Update Instructions describe the process for installation of an additional clamping mechanism provided by Siemens into the existing system, which will be activated if the gearbox is damaged. If activated, the clamping machanism will prevent the C-arm from tilting away, stop regular operation, and alert the operator to check the system and repair the gearbox damage.
- Quantity
- 131 units
- Distribution
- Nationwide distribution: States of AL, AZ, CA, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NY, OH, OK, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, and WY.
- Lot, serial or model codes
- Model numbers 10094139, 5904466, 7555365. Serial numbers 20136, 157435, 20205, 57089, 20334, 20114, 57077, 157458, 20052, 20303, 20343, 20016, 157136, 20208, 20238, 20242, 157436, 57144, 57189, 20322, 20295, 20081, 157448, 20209, 157165, 20254, 20170, 20282, 20076, 157168, 20050, 157213, 20363, 20179, 57035, 157164, 157156, 20202, 20338, 57188, 20348, 20172, 20183, 57073, 20264, 57016, 20325, 20143, 20367, 20258, 57009, 20182, 20175, 157169, 20056, 57004, 20082, 20037, 20053, 157139, 57187, 58702, 20250, 20313, 20100, 20207, 20230, 157146, 20157, 157476, 20249, 57204, 20140, 57111, 20181, 20281, 57171, 20033, 20074, 20165, 57113, 20243, 20178, 57193, 20188, 20339, 57117, 20141, 57041, 57072, 57008, 157414, 20321, 157215, 57198, 57054, 57155, 20097, 20003, 20015, 20233, 20220, 20265, 57098, 57124, 57116, 57045, 157184, 20185, 57156, 57076, 57105, 58701, 20086, 20087, 20217, 157409, 20102, 57012, 57201, 20180, 20294, 157466, 157442, 20129, 157449, 20171, 57024, 57183, 157438, and 157440.
- 510(k) numbers
- ["K010721","K021021","K073290"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28