FDA Medical Device Recalls (openFDA)
Leonhard Lang Gmbh: SKINTACT Electrodes for Defibrillation, 10X REF DF20, Rx only, Manufactured by Leonhard Lang GmbH, Archenweg 56, 6020 Innsbruck, Austria. The device is non-sterile and single use. For use on adults and children over eight years old for external defibrillation, pacing, monitoring, and cardioversion.
- Recall number
- Z-2233-2009
- Recalling firm
- Leonhard Lang Gmbh
- Product
- SKINTACT Electrodes for Defibrillation, 10X REF DF20, Rx only, Manufactured by Leonhard Lang GmbH, Archenweg 56, 6020 Innsbruck, Austria. The device is non-sterile and single use. For use on adults and children over eight years old for external defibrillation, pacing, monitoring, and cardioversion.
- Status
- Terminated
- Initiated
- 2009-07-10
- Posted
- 2009-09-24
- Terminated
- 2011-12-17
- Reason
- A potential risk was identified associated with the connectors of the defibrillation electrodes DF20. It was found that some of the connectors do not slip easily into the sockets of the defibrillator or an extension cord of the defibrillator. In such cases increased force or some wiggling is required to insert the connector fully. A potential risk for patients exist, if (a) such a connector (b)
- Root cause
- Nonconforming Material/Component
- Action
- A Leonhard Lang letter dated June 9, 2009 was sent June 6, 2009 to its customer. The letter identified the problem and the product. The letter instructed the consignee to stop using or distributing the product and to quarantine them. It also requested them to report all quantities of product that they had in stock using the attached form.
- Quantity
- 1433 USA
- Distribution
- Worldwide Distribution - US - FL Foreign: Austria, Germany, Romania, France, Dubai, Greece, Italy, South Korea, United Kingdom, Hungary, Latvia, Saudi Arabia and Lebanon
- Lot, serial or model codes
- DF20 SKINTACT Lot #'s 80506-0771, 80527-0776, 80628-0776, 80826-0773, 80911-0773,
- 510(k) numbers
- ["K072233"]
- PMA numbers
- not on file
- Product code
- MKJ
- Firm FEI
- 6736
- Firm city
- Innsbruck
- Firm state
- not on file
- Firm country
- Austria
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28