Hamilton Co: Hamilton Soft Grip Pipettes, 1 ml adjustable volume, model number 55019-40 Product is indicated for manual pipetting.
- Recall number
- Z-2233-2008
- Recalling firm
- Hamilton Co
- Product
- Hamilton Soft Grip Pipettes, 1 ml adjustable volume, model number 55019-40 Product is indicated for manual pipetting.
- Status
- Terminated
- Initiated
- 2008-05-07
- Posted
- 2008-09-17
- Terminated
- 2011-02-08
- Reason
- Manufacturing issue with spring may restrict product to drawing less than 1/2 of its stated measured capacity
- Root cause
- Device Design
- Action
- Recall initiated May 7 2008. Notification issued via certified overnight mail explains the problem and instructs customers to inventory the affected pipettes. The firm will contact its customers and supply replacement lower-body assemblies, repair instructions and return information so that affected parts can be returned to Hamilton. A fax-back form is supplied with each notification. Responses will be tracked. Contact Hamilton Co. at 1-775-558-3000 for assistance.
- Quantity
- 305 units
- Distribution
- Product distributed throughout the US and to Puerto Rico, New Zealand, Australia, Japan and Canada.
- Lot, serial or model codes
- Serial numbers 63045, 63098, 63135, 63157, 63208, 63248, 63278, 63378, 63401, 63418, 63421, 63442, 63461, 63485, 63524, 63527, 63590, 63679, 63753, 63788, 63804, 63820, 63825, 63834, 63836, 63848, 63857, 63885, 63892, 63904, 63951, 63989, 63993, 64047, 64057, 64074, 64076, 64093, 64140, 64160, 64242, 64246, 64261, 64279, 64288, 64289, 64314, 64342, 64352, 64362, 64399, 64407, 64427, 64478, 64479, 64480, 64492, 64498, 64504, 64507, 64518, 64520, 64525, 64527, 64528, 64537, 64538, 64543, 64544, 64563, 64577, 64582, 64585, 64587, 64590, 64592, 64599, 64612, 64628, 64631, 64635, 64636, 64644, 64667, 64678, 64694, 64695, 64703, 64706, 64718, 64720, 64722, 64723, 64729, 64740, 64753 64757, 64759, 64760, 64763, 64767, 64772, 64778, 64785, 64787, 64789, 64792, 64797, 64801, 64810, 65015, 65016, 65018, 65067, 65068, 65071, 65076, 65078, 65079, 65080, 65083, 65090, 65091, 65109, 65119, 65186, 65204, 65224, 65262, 65269, 65290, 65514, 65536, 65541, 65564, 65566, 65567, 65568, 65575, 65576, 65578, 65583, 65590, 65591, 65597, 65600, 65603, 65608, 65609, 65610, 65612, 65614, 65617, 65622, 65623, 65626, 65628, 65631, 65632, 65635, 65644, 65650, 65655, 65657, 65662, 65678, 65692, 65701, 65705, 65707, 65726, 65747, 65771, 65772, 65773, 65777, 65836, 65858, 65906, 65910, 65916, 65917, 65918, 65921, 65925, 65926, 65927, 65929, 65930, 65933, 65935, 65936, 65938, 65946, 65947, 65952, 65953, 65959, 65961, 65964, 65966, 65973, 65979, 65981, 65982, 65983, 65987, 65993, 65995, 65996, 65997, 65998, 65999, 66000, 66002, 66007, 66008, 66013, 66014, 66015, 66016, 66017, 66019, 66062, 66079, 66088, 66104, 66107, 66115, 66124, 66138, 66145, 66174, 66182, 66196, 66197, 66198, 66201, 66205, 66210, 66222, 66226, 66233, 66252, 66253, 66270, 66286, 66289, 66296, 66305, 66309, 66341, 66342, 66345, 66351, 66353, 66361, 66366, 66372, 66373, 66380, 66397, 66398, 66400, 66402, 66403, 66404, 66417, 66418, 66422, 66423, 66425, 66426, 66431, 66434, 66436, 66438, 66443, 66444, 66448, 66450, 66452, 66453, 66456, 66463, 66465, 66474, 66475, 66480, 66483, 66484, 66488, 66490, 66496, 66497, 66503, 66504, 66505, 66512, 66517, 66519, 66520, 66522, 66523, 66534, 66535, 66538, 66540, 66548, 66549, 66550, 66552, 66555, 66556, 66557, 66558, 66560, 66562, 66565, 66567, 66568, 66569, 66573, 66574, 66576, 66577, 66578, 66582, 66584, 66591, 66601, 66603, 66605, 66607, and 66612
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JQW
- Firm FEI
- 2915796
- Firm city
- Reno
- Firm state
- NV
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28