Mako Surgical Corporation: MAKO PN 204588 Camera 204588 Label With Camera Stand Number ***PN 200294 SN CAM XXX 2011-01.*** Weight 86.2kg.***Class I Equipment. Conforms to IEC 60601-1/A2: 1995, EN 60601-1/A2: 1995, UL 60601-1: 2003, CAN/CSA-C22.2 No. 601.1-M90.*** Manufactured in USA*** MAKO SURGICAL CORP. 2555 Davie Road, Ft. Lauderdale, FL 33317.*** 204589 Robot Arm Label***PN 203999 SN ROB XXX 2011-01.***V 100/120/230 A 9.6/8.0/4.2 Hz 50/60***Weight 39.4kg***Class I Equipment Conforms to IEC 60601-1/A2: 1995, EN 60601-1/A2: 1995, UL 60601-1:2003 CAN/CSA-C22.2 No. 601-1-M90.***Manufactured in USA***MAKO SURGICAL CORP. 2555 Davie Road, Ft. Lauderdale, FL 33317.*** 204588 Label With Guidance Module Number***PN 201251 SN GUD XXX 2011-01***Weight 51.7kg***Class I Equipment. Conforms to IEC 60601-1/A2: 1995, En 60601-1/A2: 1995, UL 60601-1: 2003, CAN/CSA-C22.2 No. 601.1-M90***Manufactured in USA***MAKO SURGICAL CORP. 2555 Davie Road, Ft. Lauderdale, FL 33317.*** THE RIO uses an infra-red camera for the purpose of spatial navigation. This camera is present on the Camera Stand and communicates with the Guidance Module to provide the locations of the tracked instruments (i.e. robotic-arm, patient trackers, probes, etc). The information is then used to provide the virtual representation of the patient anatomy and the tools on the monitor, and provides necessary information to the robotic-arm to ensure proper bone preparation.
- Recall number
- Z-2232-2011
- Recalling firm
- Mako Surgical Corporation
- Product
- MAKO PN 204588 Camera 204588 Label With Camera Stand Number ***PN 200294 SN CAM XXX 2011-01.*** Weight 86.2kg.***Class I Equipment. Conforms to IEC 60601-1/A2: 1995, EN 60601-1/A2: 1995, UL 60601-1: 2003, CAN/CSA-C22.2 No. 601.1-M90.*** Manufactured in USA*** MAKO SURGICAL CORP. 2555 Davie Road, Ft. Lauderdale, FL 33317.*** 204589 Robot Arm Label***PN 203999 SN ROB XXX 2011-01.***V 100/120/230 A 9.6/8.0/4.2 Hz 50/60***Weight 39.4kg***Class I Equipment Conforms to IEC 60601-1/A2: 1995, EN 60601-1/A2: 1995, UL 60601-1:2003 CAN/CSA-C22.2 No. 601-1-M90.***Manufactured in USA***MAKO SURGICAL CORP. 2555 Davie Road, Ft. Lauderdale, FL 33317.*** 204588 Label With Guidance Module Number***PN 201251 SN GUD XXX 2011-01***Weight 51.7kg***Class I Equipment. Conforms to IEC 60601-1/A2: 1995, En 60601-1/A2: 1995, UL 60601-1: 2003, CAN/CSA-C22.2 No. 601.1-M90***Manufactured in USA***MAKO SURGICAL CORP. 2555 Davie Road, Ft. Lauderdale, FL 33317.*** THE RIO uses an infra-red camera for the purpose of spatial navigation. This camera is present on the Camera Stand and communicates with the Guidance Module to provide the locations of the tracked instruments (i.e. robotic-arm, patient trackers, probes, etc). The information is then used to provide the virtual representation of the patient anatomy and the tools on the monitor, and provides necessary information to the robotic-arm to ensure proper bone preparation.
- Status
- Terminated
- Initiated
- 2010-10-25
- Posted
- 2011-05-20
- Terminated
- 2011-11-09
- Reason
- MAKO Surgical Corporation recalled ONE infra-red camera MAKO PN 204854 which was a component of their RIO Robotic Arm Interactive Orthopedic System with SSN P7-03127, on November 22, 2010. This component can function improperly and cause the accuracy of the system to be outside of the specification listed by the manufacturer. The recall was initiated after the camera manufacture, NDI, Waterloo, On
- Root cause
- Nonconforming Material/Component
- Action
- The firm, MAKO, sent a "RECALL NOTICE" letter dated November 22, 2010, to its customer. The letter described the product, problem and action to be taken. The customer was instructed to complete and return the attached Acknowledgement Form to their MAKOplasty Specialist (MPS) via mail to: MAKO Surgical Corp., Attn: Jim Pomeroy, 2555 Davie Road, Fort Lauderdale, FL 33317. The letter stated that "It is important that the camera be replaced immediately, and MAKO is currently addressing the situation with the camera manufacturer." In addition, MAKO will replace the camera with a camera that is not affected by the recall and verify the customers system to meet the functional specifications. If you have any questions or concerns, please contact your MPS or Sr. Director of Quality Assurance at 954-927-2044 ext 604, and /or Sr. Director, Customer Support at 954-927-2044 ext. 442.
- Quantity
- 1
- Distribution
- Nationwide distribution: NY only.
- Lot, serial or model codes
- Lot #K081867
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KQM
- Firm FEI
- 3005985723
- Firm city
- Davie
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28