Skip to the content

FDA Medical Device Recalls (openFDA)

Shent USA, Inc.: Grasper/Retriever, Alligator Jaw, Rat Tooth 2.4 mm x 8 mm x 230 cm Catalog Number: 7226 Intended use: single use accessory devices used through the working channel of a GI endoscope

Recall number
Z-2230-2021
Recalling firm
Shent USA, Inc.
Product
Grasper/Retriever, Alligator Jaw, Rat Tooth 2.4 mm x 8 mm x 230 cm Catalog Number: 7226 Intended use: single use accessory devices used through the working channel of a GI endoscope
Status
Completed
Initiated
2021-01-12
Posted
not on file
Terminated
not on file
Reason
Mislabeled: Devices labeled as catalog #7208 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 230 cm) may actually be catalog #7226 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 8 mm x 230 cm) and Devices labeled as catalog #7226, (Grasper/Retriever, Alligator Jaw, Rat Tooth, 2.4 mm x 8 mm x 230 cm) may actually be catalog #7208, Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 230 cm).
Root cause
Packaging process control
Action
Telemed issued Urgent Medical Device Recall letter on 1/12/21. Letter states reason for recall, health risk and action to take: Please respond to this recall notification (email response is acceptable) to let us know whether or not you have any of the recalled product in your inventory. If you are in possession of any recalled devices, please provide the quantity, catalog number and lot number. All affected devices must be returned to TeleMed; upon receiving your response to this notification, TeleMed Customer Service will provide you with an RGA number for product return. If you have any questions regarding this medical device recall, please contact: prakash@telemedsystems.com. 508-723-6609; Customerservice@telemedsystems.com 978-567-9033
Quantity
85 units OUS
Distribution
US Nationwide Distribution - CA, CO, MI Foreign: Australia
Lot, serial or model codes
Lot Code: D20177226-19
510(k) numbers
["K974300"]
PMA numbers
not on file
Product code
OCZ
Firm FEI
1222168
Firm city
Hudson
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register