Skip to the content

FDA Medical Device Recalls (openFDA)

ArjoHuntleigh: Maxi Move Scoop Stretcher; Note: The stretcher is an accessory to the Maxi Move Patient Lift; Made by Arjo Hospital Equipment AB, Eslov, Sweden; product number KRA2210-20.

Recall number
Z-2230-2009
Recalling firm
ArjoHuntleigh
Product
Maxi Move Scoop Stretcher; Note: The stretcher is an accessory to the Maxi Move Patient Lift; Made by Arjo Hospital Equipment AB, Eslov, Sweden; product number KRA2210-20.
Status
Terminated
Initiated
2009-08-18
Posted
2009-09-22
Terminated
2010-01-07
Reason
A Scoop Stretcher is made up of two frames that are held together with two locking devices, one on each end. Each locking device contains a safety latch. The safety latch could unintentionally release when the side frames are pulled for positioning.
Root cause
Manufacturing material removal
Action
ArjoHuntleigh sent their customers a cover letter and the Field Safety Notification, informing them of the potential problem with the safety latches on the stretcher and a pending visit from an ArjoHuntleigh Service Technician to correct the problem. These letters were sent via UPS 2nd day delivery on 8/18/09. The customers were requested to test their stretchers upon receipt of the letter following the instructions for checking the safety latches on the stretcher: * Place the stretcher on its side on a non-scratch surface. * Try to pull the two sides apart with the safety latch engaged. * Test both the head and foot ends of the stretcher. * If the stretcher pulls apart, remove it from use immediately. It is important to note that all affected stretchers will be upgraded, whether they fail the test or not.
Quantity
214 units
Distribution
Worldwide Distribution
Lot, serial or model codes
all six digit serial numbers beginning with 6xxxxx and all serial numbers 08xxxxxxxx up to 0812001186.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSA
Firm FEI
3007515958
Firm city
Addison
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register