FDA Medical Device Recalls (openFDA)
ArjoHuntleigh: Maxi Move Scoop Stretcher; Note: The stretcher is an accessory to the Maxi Move Patient Lift; Made by Arjo Hospital Equipment AB, Eslov, Sweden; product number KRA2210-20.
- Recall number
- Z-2230-2009
- Recalling firm
- ArjoHuntleigh
- Product
- Maxi Move Scoop Stretcher; Note: The stretcher is an accessory to the Maxi Move Patient Lift; Made by Arjo Hospital Equipment AB, Eslov, Sweden; product number KRA2210-20.
- Status
- Terminated
- Initiated
- 2009-08-18
- Posted
- 2009-09-22
- Terminated
- 2010-01-07
- Reason
- A Scoop Stretcher is made up of two frames that are held together with two locking devices, one on each end. Each locking device contains a safety latch. The safety latch could unintentionally release when the side frames are pulled for positioning.
- Root cause
- Manufacturing material removal
- Action
- ArjoHuntleigh sent their customers a cover letter and the Field Safety Notification, informing them of the potential problem with the safety latches on the stretcher and a pending visit from an ArjoHuntleigh Service Technician to correct the problem. These letters were sent via UPS 2nd day delivery on 8/18/09. The customers were requested to test their stretchers upon receipt of the letter following the instructions for checking the safety latches on the stretcher: * Place the stretcher on its side on a non-scratch surface. * Try to pull the two sides apart with the safety latch engaged. * Test both the head and foot ends of the stretcher. * If the stretcher pulls apart, remove it from use immediately. It is important to note that all affected stretchers will be upgraded, whether they fail the test or not.
- Quantity
- 214 units
- Distribution
- Worldwide Distribution
- Lot, serial or model codes
- all six digit serial numbers beginning with 6xxxxx and all serial numbers 08xxxxxxxx up to 0812001186.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FSA
- Firm FEI
- 3007515958
- Firm city
- Addison
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28