FDA Medical Device Recalls (openFDA)

St. Jude Medical: St. Jude Medical FAST-CATH Hemostasis Introducer 4F, REF 406709, Length 12 CM, Lot 4219 Max. Guidewire O.D. .038", 2010-01, Manufacture, St. Jude Medical, 14901 DeVeau place, Minnetonka, MN 55345-2126, Sterile EO. Used for the introduction of catheters into a vessel.

Recall number
Z-2229-2009
Recalling firm
St. Jude Medical
Product
St. Jude Medical FAST-CATH Hemostasis Introducer 4F, REF 406709, Length 12 CM, Lot 4219 Max. Guidewire O.D. .038", 2010-01, Manufacture, St. Jude Medical, 14901 DeVeau place, Minnetonka, MN 55345-2126, Sterile EO. Used for the introduction of catheters into a vessel.
Status
Terminated
Initiated
2009-08-17
Posted
2009-09-18
Terminated
2011-12-17
Reason
St. Jude Medical has initiated a voluntary recall of two lots of Fast-Cath Hemostasis Introducers-reorder number 406709, lot number 4219 and reorder number 406702, lot number 6372. They discovered that the product lots could contain some items which have been mis-labeled with the incorrect French size. In one case, the pouch label does not correctly identify the content of the pouch. In another case, the carton label is incorrect. The pouch label indicated that the product was 4F catheter, but the contents of the pouch may be a 7F catheter. St. Jude Medical has not received any reports of injuries resulting from this mis-labeling.
Root cause
Error in labeling
Action
Field Representatives were sent a "St Jude Medical Urgent Product Information" letter to be hand delivered to their customers. The letter listed the products and problem and advised consignees to discontinue use of the product. A form was attached for the return of products to the firm. Contact your local St Jude Medical representative for further information.
Quantity
490
Distribution
Worldwide Distribution -- US (CA, MI, PA and TN), China and Belgium.
Lot, serial or model codes
Lot Number: 4219
510(k) numbers
["K791129"]
PMA numbers
not on file
Product code
DQX
Firm FEI
2182269
Firm city
Minnetonka
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register