ConvaTec, Inc: Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;

Recall number
Z-2228-2017
Recalling firm
ConvaTec, Inc
Product
Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;
Status
Terminated
Initiated
2017-05-10
Posted
2017-05-24
Terminated
2021-08-27
Reason
Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the oxygen tubing and the oxygen mask connector. While the defect affects less than 2% of recalled product, there is a potential for harm when oxygen supply to the user is disrupted.
Root cause
Process control
Action
Recall Letters were sent to consignees on May 10, 2017. The letter requested that they immediately stop distribution and quarantine all affected lots. The letter also requested a sub-recall if the product had been further distributed. Also attached to the letter is a Recall Response Form to be returned.
Quantity
816,709
Distribution
CA, CT, FL, GA, IL, MO, NJ, NY, NC, OH, PA, RI, VA , Puerto Rico, Australia, Belgium, Canada, Columbia, Czech, Denmark, Finland, France, Germany, Greece, Israel, Italy, Japan, Jordan, Malaysia, Malta, Netherlands, Norway, Rwanda, Saudi Arabia, Sweden, Switzerland, United Kingdom, and Vietnam
Lot, serial or model codes
Lot Number 102153, 102339, 102941, 103176, 103453, 103718, 103919, 104471, 105034, 105282, 105427, 105567, 105996, 106960, 108658, 112024, 112530
510(k) numbers
not on file
PMA numbers
not on file
Product code
BYG
Firm FEI
3011987967
Firm city
Greensboro
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register