Smith & Nephew, Inc. Endoscopy Division: OSTEORAPTOR CURVED 2.3 mm Suture Anchor SA ULTRABRAID COBRAID Suture, Blue; Part Number: 72203290 Biodegradable suture anchor

Recall number
Z-2228-2013
Recalling firm
Smith & Nephew, Inc. Endoscopy Division
Product
OSTEORAPTOR CURVED 2.3 mm Suture Anchor SA ULTRABRAID COBRAID Suture, Blue; Part Number: 72203290 Biodegradable suture anchor
Status
Terminated
Initiated
2013-08-06
Posted
2013-09-17
Terminated
2016-04-20
Reason
Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
Root cause
Packaging
Action
Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.
Quantity
10296 US
Distribution
Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.
Lot, serial or model codes
50364854 50366872 50366972 50366980 50366985 50366989 50370760 50370771 50370940 50371292 50376475 50385813 50386237 50386563 50386749 50388066 50391012 50394695 50396184 50400301 50400306 50400307 50403488 50404241 50406445 50407524 50407571 50410789 50411371 50414977 50415258 50415670 50416148 50417206 50418093 50418993 50420658 50423443 50429652 50430190 50430402 50431731 50433933 50435319 50441441 50442649 50444415 50445034 50445760 50447482 50448060 50448722 50450469 50451783 50453148 50454104 50455627 50456912 50458608 50460242
510(k) numbers
["K082215"]
PMA numbers
not on file
Product code
JDR
Firm FEI
3003604053
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register