FDA Medical Device Recalls (openFDA)

Philips Respironics, Inc.: DreamStation Auto CPAP. Non-Continuous Ventilator.

Recall number
Z-2227-2025
Recalling firm
Philips Respironics, Inc.
Product
DreamStation Auto CPAP. Non-Continuous Ventilator.
Status
Open, Classified
Initiated
2025-06-30
Posted
2025-08-08
Terminated
not on file
Reason
Devices may possess a programming error resulting in an incorrect device configuration.
Root cause
Process control
Action
Consignees were contacted by phone from 6/30/25 to 7/02/25 to facilitate provision of a replacement device. Additionally, a "Medical Device Recall Field Safety Notice" dated 6/6/2025 was sent to consignees instructing them to continue using their affected device until Philips Respironics can ship them a replacement. DMEs are to share the notice with those who need to be aware within their organization and to organizations where devices were transferred. If a consignee received the written notice but not a phone call from Philips Respironics, they are to call 1-877-387-3311 or email PatientSupport@philips.com. Consignees with any questions can contact Philips Respironics at 1-800-345-6443 (prompts 4, 5).
Quantity
57 units
Distribution
Worldwide - US Nationwide in the states of AL, CA, CO, FL, GA, IN, LA, MD, NE, NJ, NY, PA, TX & DC. The country of France.
Lot, serial or model codes
Model No. UDSX500S11F; UDI: 606959429765; Serial No. J204432084D03, J206957368810, J207023218478, J21450857096A, J219786492F58, J222909239F3E, J22718022B9D6, J22924665CF8A, J230303131D07, J23148826984B, J23307132CC16, J2379520271A4, J23933071851F, J23963189F617, J242941498857, J243763392C02, J24551698A015, J247575014664, J24858283A8C3, J24994323EB40, J2529261301FF, J2543728920D9, J254401628C83, J25467463E51E, J25493556DA17, J254990254097, J255312624BAB, J256278994969, J25782569130A, J257828377BB3, J260977942F33, J2617246993BE, J26312750F8BD, J263183345586, J26751119CCB4, J27018735B4C6, J270987383503, J271152226FB8, J27220570C1F1, J27464550B547, J275917117696, J277824942640, J282846981F7D, J28582229DE76, J2901054169AB, J2902459395E1, J2921836070C5, J29746839268B, J2982204155F4, J299187060F67, J29925204838B, J299567590A7A, J2997189741CA, J30005981C03D, J301116505688, J3067824786A4, J30982199BB1D.
510(k) numbers
["K131982"]
PMA numbers
not on file
Product code
BZD
Firm FEI
2518422
Firm city
Murrysville
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register