FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: DYND3000xxP Series - Flexible Laryngeal Mask Airway Packs

Recall number
Z-2227-2023
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
DYND3000xxP Series - Flexible Laryngeal Mask Airway Packs
Status
Open, Classified
Initiated
2023-03-28
Posted
2023-07-21
Terminated
not on file
Reason
The case and each (i.e. packet) labels for the DYND3000xxP series do not include an expiration date, however, the lubricating jelly (MDS032280) included in the DYNDP3000xxP series is labeled with an expiration date.
Root cause
Labeling Change Control
Action
The consignee recall notification was sent out via first class mail and email on 03/28/2023. The letter instructs the consignee to over-label the Flexible Laryngeal Mask Airway Packs stating to Check the Expiration date of Lubricating Jelly MDS032280 and do not use if Expired . No product is to be returned back to Medline Industries, LP.
Quantity
9945 ea
Distribution
Worldwide distribution - US Nationwide and the country of Bermuda.
Lot, serial or model codes
All Lots for series since product launch; Model Number (Unique Device Identifier) -DYND300010P (Each level packaging: 10193489010671; Case level packaging: 20193489010678); DYND300015P (Each level packaging: 10193489010688; Case level packaging: 20193489010685); DYND300020P (Each level packaging: 10193489010695; Case level packaging: 20193489010692); DYND300025P (Each level packaging: 10193489010701 Case level packaging: 20193489010708); DYND300030P (Each level packaging: 10193489010718; Case level packaging: 20193489010715); DYND300040P (Each level packaging: 10193489010725; Case level packaging: 20193489010722); DYND300050P (Each level packaging: 10193489010732; Case level packaging: 20193489010739)
510(k) numbers
not on file
PMA numbers
not on file
Product code
CAE
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register