Skip to the content

FDA Medical Device Recalls (openFDA)

Orthosoft, Inc. dba Zimmer CAS: Sesamoid Plasty, Stereotaxic Instrument The Sesamoid Plasty workstation is used to run the tracking software as part of the software based navigation systems. The affected Sesamoid Plasty components are the central structural column and its base that hold up the computer and the tracking camera on the Sesamoid Plasty workstation. They are also designed to be disassembled such that the column can be folded onto itself for ease of transport after use. This is a retrospective recall report for a removal initiated on 17 June 2008

Recall number
Z-2227-2017
Recalling firm
Orthosoft, Inc. dba Zimmer CAS
Product
Sesamoid Plasty, Stereotaxic Instrument The Sesamoid Plasty workstation is used to run the tracking software as part of the software based navigation systems. The affected Sesamoid Plasty components are the central structural column and its base that hold up the computer and the tracking camera on the Sesamoid Plasty workstation. They are also designed to be disassembled such that the column can be folded onto itself for ease of transport after use. This is a retrospective recall report for a removal initiated on 17 June 2008
Status
Terminated
Initiated
2008-06-04
Posted
not on file
Terminated
2017-06-02
Reason
Zimmer CAS voluntarily conducted a retrospective recall of the Sesamoid Plasty due to the risk of the top of the workstation to potentially fall over during disassembly.
Root cause
Device Design
Action
Zimmer CAS conducted a retrospective recall for the Sesamoid Plasty CAS workstation due to the possible risk of injury if the workstation is not properly disassembled. The firm mailed notices dated June 4, 2008. The original Sesamoid Plasty design did not include a safety design feature which is now available on current Sesamoid Plasty Workstations. The Safety design feature is a mechanism that prevents a latch lock at the base of the workstation column from being opened to allow folding of the workstation column for transport after use unless the camera and its arm are first removed from the top of the workstation. A safety label was initially used instead that instructed the user of the recommended disassembly sequence (i.e. camera and arm first). Once consignees receive their replacement system, they were instructed to contact Zimmer CAS Customer Service to assist them with the return (RMA) of the Sesamoid Plasty Workstation. For any questions or concerns please contact the Zimmer CAS customer service, 1-866-336-7846.
Quantity
10
Distribution
Worldwide Distribution - US including FL, CA Foreign: Austrailia, New Zealand, France, Brazil
Lot, serial or model codes
Part Number and : 521.025/521.028; Lot Number/Serial Numbers: SP-001, SP-002, SP-003, SP-004, SP-005, SP-006,  SP-007, SP-008, SP-009, SP-010
510(k) numbers
["K071929"]
PMA numbers
not on file
Product code
HAW
Firm FEI
3003714383
Firm city
Montreal
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register