Skip to the content

FDA Medical Device Recalls (openFDA)

Biomerieux Inc: ETEST IMIPENEM RELEBACTAM. in vitro diagnostic

Recall number
Z-2226-2021
Recalling firm
Biomerieux Inc
Product
ETEST IMIPENEM RELEBACTAM. in vitro diagnostic
Status
Terminated
Initiated
2021-06-24
Posted
not on file
Terminated
2025-02-10
Reason
Major errors (Resistant result instead of Susceptible result) were observed with Pseudomonas aeruginosa strains.
Root cause
Under Investigation by firm
Action
The recall notification was issued to the impacted subsidiaries and distributors on 24JUN2021. The notification includes a Customer Letter to be translated into local language and sent to each impacted customer. The customer letter will inform the customers about the MIC overestimation and false resistant results issue and will be instructed to: - Perform a retest using an alternative method on P. aeruginosa strains for which they obtained resistant or intermediate results with ETEST IMIPENEM RELEBACTAM ref. 420925 and 420927 - Among tests previously performed, identify any possible false resistant/intermediate results that may have occurred, to analyze the related risks and to determine appropriate actions if relevant.
Quantity
1,431
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AR, AZ, CA, CO, DE, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii. The countries of Austria, Belgium, Croatia, Czech Republic, France, Germany, Greece, Italy, Netherlands, Poland, Portugal, Spain, Sweden, Switzerland and United Kingdom.
Lot, serial or model codes
Ref 420927 (US Product) Lot numbers 1007888860, 1008598340, 1008701580, 1007607290, 1007888870, 1008390030, 1008598350 and 1008701590 and  Ref 420925 (not distributed in US):Lot Codes: 1007888860, 100859840 and 1008701580
510(k) numbers
["K191953"]
PMA numbers
not on file
Product code
JWY
Firm FEI
1950204
Firm city
Hazelwood
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register