Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc. Endoscopy Division: OSTEORAPTOR 2.9 mm Suture Anchor with 1 ULTRABRAID Suture, White; Part Number: 72202165 Biodegradable suture anchor

Recall number
Z-2226-2013
Recalling firm
Smith & Nephew, Inc. Endoscopy Division
Product
OSTEORAPTOR 2.9 mm Suture Anchor with 1 ULTRABRAID Suture, White; Part Number: 72202165 Biodegradable suture anchor
Status
Terminated
Initiated
2013-08-06
Posted
2013-09-17
Terminated
2016-04-20
Reason
Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
Root cause
Packaging
Action
Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.
Quantity
4452 US
Distribution
Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.
Lot, serial or model codes
50272335 50272468 50274466 50274792 50275817 50278112 50282209 50283021 50286518 50286657 50286925 50288725 50290551 50291360 50294564 50295956 50296887 50298073 50301797 50307182 50309401 50309926 50310781 50312498 50314440 50319909 50320249 50322931 50323640 50325599 50326068 50329810 50329975 50333958 50335745 50338377 50341064
510(k) numbers
["K082215"]
PMA numbers
not on file
Product code
JDR
Firm FEI
3003604053
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register