Beckman Coulter Inc.: SYNCHRON Systems Calibrator 1, Part Numbers: 468405, Lots prior to M005558 The Beckman Coulter SYNCHRON Calibrator 1 (CAL 1), used in conjunction with SYNCHRON reagents, is intended for the calibration of immunoprotein tests on SYNCHRON LX and UniCel DxC Systems.

Recall number
Z-2226-2011
Recalling firm
Beckman Coulter Inc.
Product
SYNCHRON Systems Calibrator 1, Part Numbers: 468405, Lots prior to M005558 The Beckman Coulter SYNCHRON Calibrator 1 (CAL 1), used in conjunction with SYNCHRON reagents, is intended for the calibration of immunoprotein tests on SYNCHRON LX and UniCel DxC Systems.
Status
Terminated
Initiated
2011-03-11
Posted
2011-05-20
Terminated
2012-05-04
Reason
The recall was initiated because standardization of the Synchron Systems Calibrator 1, used for both the Synchron LX and UniCel DxC systems, has been established to the new IFCC Reference Preparation for Plasma Proteins, ERM-DA470k (formerly CR470) beginning with lot number M005558. During the standardization to the new IFCC standard, ERM-DA470k, we confirmed that Complement C4 (C4) had been re
Root cause
Pending
Action
Beckman Coulter initiated a Product Corrective Action (PCA) letter on the week of March 14, 2011, to the affected customers. Though Beckman is recalling the LX Cal 1, the recall letter was sent to the C4 reagent customers, as these are the only customers who will be impacted by the problem noted in the PCA letter. The letter provided the customers with an explanation of the problem identified and an action to be taken. The letter recommended consignees take the following actions: (1)Notify their clients to review and evaluate the impact on patients whose Complement C4 recovered near the clinical decision points. (2) Re-evaluate their laboratory's reference ranges for C4. The letter further requested that consignees share this information with laboratory staff and retain this notification as part of your laboratory Quality System documentation. Questions regarding this Product Announcement letter, or requests for the correlation graphs should be directed to Customer Support Center (Call Center) on the Beckman Coulter website at www.BeckmanCoulter.com/Customer Support/Support Request or call (800) 854-3633 in the United States and Canada.
Quantity
96,173 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Belgium, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Czech Republic, Denmark, Eritrea, France, French Polynesia, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Republic of Kuwait, Lebanon, Lithuania, Malaysia, Mexico, Namibia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Poland, Puerto Rico, Qatar Reunion, Russian Federation, Rwanda, Serbia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, and Viet Nam
Lot, serial or model codes
Lots prior to M005558
510(k) numbers
["K965108"]
PMA numbers
not on file
Product code
JIT
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register