FDA Medical Device Recalls (openFDA)
Applied Medical Resources Corp: Applied Medical 15x100mm Kii Optical Access System, Non-Threaded 6/Box, Model Number: C0R36, Labeled Sterile. Disposable single-use devices indicated for use in general, abdominal, gynecological and thoracic minimally invasive surgeries. Endoscopic instruments for surgery.
- Recall number
- Z-2226-2009
- Recalling firm
- Applied Medical Resources Corp
- Product
- Applied Medical 15x100mm Kii Optical Access System, Non-Threaded 6/Box, Model Number: C0R36, Labeled Sterile. Disposable single-use devices indicated for use in general, abdominal, gynecological and thoracic minimally invasive surgeries. Endoscopic instruments for surgery.
- Status
- Terminated
- Initiated
- 2009-06-04
- Posted
- 2009-09-18
- Terminated
- 2011-07-14
- Reason
- Applied Medical is conducting a coluntary recall of two lot numbers, 1084426 and 1085506, of the 15mm Kii non-threaded optical access system, model C0R36, due to a potential weakness in the cannula tip, possibly resulting in tip breakage.
- Root cause
- Component design/selection
- Action
- Applied Medical Resources Corp. notified Customers via an overnight "Voluntary Recall Notification" Letter dated June 9, 2009. The letter informed customers of the reason for recall and asked that the listed product be returned. The recall letter provided customers with a contact and telephone number for product return questions (1-949-713-8652) and a separate contact for regulatory questions (1-949-713-8369).
- Quantity
- 934 units
- Distribution
- Class II Recall - Worldwide -- United States, Australia, New Zealand, France, Canada, Norway and Great Britain.
- Lot, serial or model codes
- Model Number: C0R36, Lot Numbers:1084426 and 1085506.
- 510(k) numbers
- ["K060096"]
- PMA numbers
- not on file
- Product code
- GCJ
- Firm FEI
- 2027111
- Firm city
- Rancho Santa Margarita
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28