Magellan Diagnostics, Inc.: Magellan Diagnostics LeadCare and LeadCare II test Kits. Product Usage: The LeadCare Blood Lead Test Kit is for in vitro diagnostic use only. The Test Kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and Treatment Reagent of the Test Kit are specific for lead only. Contents of this LeadCare Blood Lead Test Kit must only be used with a LeadCare Analyzer and Blood Lead Testing System.

Recall number
Z-2225-2017
Recalling firm
Magellan Diagnostics, Inc.
Product
Magellan Diagnostics LeadCare and LeadCare II test Kits. Product Usage: The LeadCare Blood Lead Test Kit is for in vitro diagnostic use only. The Test Kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and Treatment Reagent of the Test Kit are specific for lead only. Contents of this LeadCare Blood Lead Test Kit must only be used with a LeadCare Analyzer and Blood Lead Testing System.
Status
Terminated
Initiated
2017-05-23
Posted
2017-06-05
Terminated
2021-07-08
Reason
Underestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems
Root cause
Device Design
Action
Magellan issued a Safety Notification letter from May 23-25, 2017 . The firm states following the FDA Safety Notice, published May 17, 2017, this notification is to advise your facility regarding a change to existing LeadCare product usage and labeling. Prior instructions for use included capillary and venous blood samples. However, because the LeadCare Testing Systems may underestimate blood lead levels and give inaccurate results when processing venous blood samples, the FDA recommends discontinuing the use of venous blood samples with any LeadCare Blood Lead Testing System (LeadCare, LeadCare II, LeadCare Ultra, LeadCare Plus). Capillary samples may still be used on all LeadCare prducts listed above. Users requested to complete/sign Fax Form Record to verify that you have read and understood the Safety Communication to Customers on page 1. Contact Reba Daoust, Director of QA/RA 978-248-4811 or rdaoust@magellandx.com
Quantity
171,979 kits (total)= 153,447 (US)and 18532 (OUS)
Distribution
Worldwide Distribution - US (Nationwide) Foreign: AUSTRALIA AUSTRIA BOLIVIA CANADA CHILE CHINA COLOMBIA DOMINICAN REPUB ENGLAND GERMANY INDIA UK ISRAEL ITALY JAPAN KOREA MALI MYANMAR SPAIN NEPAL NETHERLANDS NEW ZEALAND P.R. CHINA PANAMA US PERU PHILLIPINES POLAND PORTUGAL REP OF KOSOVO SOUTH AFRICA SRI LANKA USA SWEDEN SWITZERLAND TRINIDAD UNITED ARAB EMI VIETNAM
Lot, serial or model codes
All lot codes
510(k) numbers
["K971640"]
PMA numbers
not on file
Product code
DOF
Firm FEI
1218996
Firm city
North Billerica
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register