Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc. Endoscopy Division: OSTEORAPTOR 2.9 mm Suture Anchor with 2 ULTRABRAID Suture, White / Black; Part Number: 72201996 Biodegradable suture anchor

Recall number
Z-2225-2013
Recalling firm
Smith & Nephew, Inc. Endoscopy Division
Product
OSTEORAPTOR 2.9 mm Suture Anchor with 2 ULTRABRAID Suture, White / Black; Part Number: 72201996 Biodegradable suture anchor
Status
Terminated
Initiated
2013-08-06
Posted
2013-09-17
Terminated
2016-04-20
Reason
Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
Root cause
Packaging
Action
Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.
Quantity
5873 US
Distribution
Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.
Lot, serial or model codes
50272330 50272467 50274209 50274465 50274791 50275815 50277800 50292667 50296071 50302316 50306994 50310605 50311297 50316179 50318681 50321243 50323086 50325048 50325601 50326880 50329010 50329811 50334408 50336660 50338997 50340629 50344015 50346092 50346624 50353156 50354524 50356989 50358564 50363890 50364919 50368628 50370206 50374882 50376601 50378711 50381816 50385513 50390999 50391628 50392410 50395539 50400939 50403274 50404746 50406677 50408208 50409818 50412886 50414333 50415150 50416320 50417391 50418417 50419567 50421499 50431562 50433945 50436052 50437097 50439085 50445036 50445557 50447140 50449470 50450754 50452349 50453811 50454844 50456056 50458436 50461697
510(k) numbers
["K082215"]
PMA numbers
not on file
Product code
JDR
Firm FEI
3003604053
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register