Fresenius Medical Care North America: Fresenius 2008K, 2008K2 and 2008 K2@home Hemodialysis Machines, Manufactured by Fresenius USA Inc., Walnut Creek, CA
- Recall number
- Z-2225-2010
- Recalling firm
- Fresenius Medical Care North America
- Product
- Fresenius 2008K, 2008K2 and 2008 K2@home Hemodialysis Machines, Manufactured by Fresenius USA Inc., Walnut Creek, CA
- Status
- Terminated
- Initiated
- 2009-11-20
- Posted
- 2010-08-15
- Terminated
- 2011-02-07
- Reason
- Power cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy.
- Root cause
- Component design/selection
- Action
- An urgent device recall notification was sent to all consignees. Customers are asked to inspect cords for defects, and are refered to preventive maintenance requirements in the manual. Customers who identify the defective cords will be instructed to discontinue use as soon as possible without jeopardizing patient care. Customers are instructed to monitor cords regularly, and pay attention to machines that are frequently plugged or unplugged. Personnel are instructed to not use the cord to unplug the device from the wall (yanking the plug from the wall) and to refrain from rolling over the cord or plug.
- Quantity
- 60885
- Distribution
- Product distributed to 5404 customers throughout the US and the world.
- Lot, serial or model codes
- Catalog numbers: 2008K: 190300, 190303, 190305, 190336, 190371, 190372, 190373, 190517, R190305 (Refurbished), R190371 (Refurbished), RTL190371, RTL190373, RTLR190305 (Refurbished), RTLR190371 (Refurbished), RTLR190373 (Refurbished)
- 510(k) numbers
- ["K994267"]
- PMA numbers
- not on file
- Product code
- KDI
- Firm FEI
- 3001451489
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28