Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corp.: REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. D684-00-0403-05 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. D684-00-0403-06 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. D684-00-0403-10

Recall number
Z-2224-2023
Recalling firm
Datascope Corp.
Product
REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. D684-00-0403-05 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. D684-00-0403-06 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. D684-00-0403-10
Status
Open, Classified
Initiated
2023-06-23
Posted
2023-07-21
Terminated
not on file
Reason
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Root cause
Under Investigation by firm
Action
Firm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and is working with the supplier towards a resolution. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Customer Service at 1-888-943-8872, options 2, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Quantity
321,609 total kits
Distribution
Domestic distribution nationwide. International distribution worldwide.
Lot, serial or model codes
Product Description UDI REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. 10607567106656 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. 10607567106694 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. 10607567107943
510(k) numbers
["K002365"]
PMA numbers
not on file
Product code
DYB
Firm FEI
1000222374
Firm city
Fairfield
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register