Skip to the content

FDA Medical Device Recalls (openFDA)

Osteotec Limited: Brand Name: Osteotec Product Name: Osteotec Silicone Finger Implant Model/Catalog Number: OSTF-3 (Sz. 3), OSTF-4 (Sz. 4), OSTF-5 (Sz. 5) Software Version: N/A Product Description: The Osteotec Silicone Finger Implant is a one-piece, sterile, flexible silicone elastomer implant which is placed, during arthroplasty, in the proximal and distal intramedullary canals of either metacarpophalangeal (MCP) or proximal interphalangeal (PIP) joints. The implant is not fixed into the bones and becomes stabilized by the encapsulation process. The implant is available in eleven sizes to meet various anatomical variations and is supplied sterile. Component: N/A

Recall number
Z-2222-2025
Recalling firm
Osteotec Limited
Product
Brand Name: Osteotec Product Name: Osteotec Silicone Finger Implant Model/Catalog Number: OSTF-3 (Sz. 3), OSTF-4 (Sz. 4), OSTF-5 (Sz. 5) Software Version: N/A Product Description: The Osteotec Silicone Finger Implant is a one-piece, sterile, flexible silicone elastomer implant which is placed, during arthroplasty, in the proximal and distal intramedullary canals of either metacarpophalangeal (MCP) or proximal interphalangeal (PIP) joints. The implant is not fixed into the bones and becomes stabilized by the encapsulation process. The implant is available in eleven sizes to meet various anatomical variations and is supplied sterile. Component: N/A
Status
Open, Classified
Initiated
2025-06-18
Posted
2025-07-31
Terminated
not on file
Reason
There is the potential that the silicone implant may contain foreign material
Root cause
Process change control
Action
On June 18, 2025, Osteotec issued an Urgent Medical Device Recall Notification via E-Mail. Osteotec asked consignees to take the following actions: 1. Immediately review your inventory and quarantine any of the affected LOTs. 2. If a physician has successfully implanted the device, they physician should continue to monitor the implant for any infections or issues. If a physician has successfully implanted a device that potentially could contain foreign material, it is recommended that they continue to monitor the patient for any infections or issues associated with the recalled device. 3. Share this Product Recall letter within your facility network and forward to any customers to whom you have distributed products from the affected LOTs to ensure awareness. 4. Complete and return the attached response and acknowledgement form to Osteotec. 5. Return affected products to your Agent/Distributor/Osteotec
Quantity
15573 units
Distribution
US Nationwide distribution in the state of CA.
Lot, serial or model codes
Lot Code: OSTF-3 LOT 127707 Exp 07/17/2028, UDI-DI:5060183090040 OSTF-4 LOT 127708 Exp 07/17/2028, UDI-DI:5060183090057 OSTF-5 LOT 127709 Exp 07/17/2028, UDI-DI:5060183090064
510(k) numbers
["K140453"]
PMA numbers
not on file
Product code
KYJ
Firm FEI
3008395366
Firm city
Christchurch
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register