FDA Medical Device Recalls (openFDA)
Osteotec Limited: Brand Name: Osteotec Product Name: Osteotec Silicone Finger Implant Model/Catalog Number: OSTF-3 (Sz. 3), OSTF-4 (Sz. 4), OSTF-5 (Sz. 5) Software Version: N/A Product Description: The Osteotec Silicone Finger Implant is a one-piece, sterile, flexible silicone elastomer implant which is placed, during arthroplasty, in the proximal and distal intramedullary canals of either metacarpophalangeal (MCP) or proximal interphalangeal (PIP) joints. The implant is not fixed into the bones and becomes stabilized by the encapsulation process. The implant is available in eleven sizes to meet various anatomical variations and is supplied sterile. Component: N/A
- Recall number
- Z-2222-2025
- Recalling firm
- Osteotec Limited
- Product
- Brand Name: Osteotec Product Name: Osteotec Silicone Finger Implant Model/Catalog Number: OSTF-3 (Sz. 3), OSTF-4 (Sz. 4), OSTF-5 (Sz. 5) Software Version: N/A Product Description: The Osteotec Silicone Finger Implant is a one-piece, sterile, flexible silicone elastomer implant which is placed, during arthroplasty, in the proximal and distal intramedullary canals of either metacarpophalangeal (MCP) or proximal interphalangeal (PIP) joints. The implant is not fixed into the bones and becomes stabilized by the encapsulation process. The implant is available in eleven sizes to meet various anatomical variations and is supplied sterile. Component: N/A
- Status
- Open, Classified
- Initiated
- 2025-06-18
- Posted
- 2025-07-31
- Terminated
- not on file
- Reason
- There is the potential that the silicone implant may contain foreign material
- Root cause
- Process change control
- Action
- On June 18, 2025, Osteotec issued an Urgent Medical Device Recall Notification via E-Mail. Osteotec asked consignees to take the following actions: 1. Immediately review your inventory and quarantine any of the affected LOTs. 2. If a physician has successfully implanted the device, they physician should continue to monitor the implant for any infections or issues. If a physician has successfully implanted a device that potentially could contain foreign material, it is recommended that they continue to monitor the patient for any infections or issues associated with the recalled device. 3. Share this Product Recall letter within your facility network and forward to any customers to whom you have distributed products from the affected LOTs to ensure awareness. 4. Complete and return the attached response and acknowledgement form to Osteotec. 5. Return affected products to your Agent/Distributor/Osteotec
- Quantity
- 15573 units
- Distribution
- US Nationwide distribution in the state of CA.
- Lot, serial or model codes
- Lot Code: OSTF-3 LOT 127707 Exp 07/17/2028, UDI-DI:5060183090040 OSTF-4 LOT 127708 Exp 07/17/2028, UDI-DI:5060183090057 OSTF-5 LOT 127709 Exp 07/17/2028, UDI-DI:5060183090064
- 510(k) numbers
- ["K140453"]
- PMA numbers
- not on file
- Product code
- KYJ
- Firm FEI
- 3008395366
- Firm city
- Christchurch
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28