GE OEC Medical Systems, Inc: GEHC OEC 8800 Mobile Fluoroscopic X-ray System, GE Healthcare, Surgery, Salt Lake City, UT. Mobile fluoroscopic examination of human anatomy.

Recall number
Z-2221-2012
Recalling firm
GE OEC Medical Systems, Inc
Product
GEHC OEC 8800 Mobile Fluoroscopic X-ray System, GE Healthcare, Surgery, Salt Lake City, UT. Mobile fluoroscopic examination of human anatomy.
Status
Terminated
Initiated
2007-10-19
Posted
2012-08-17
Terminated
2012-08-17
Reason
X-ray units did not meet manufacturer's specification for linearity.
Root cause
Nonconforming Material/Component
Action
GE updated their user manual and changed their internal specifications to match those of the CFR In addition, a software code correction was to be made to forward productions in order to correct the reported defect. *** This recall issued was previously classified under an incorrect recall number. It is being reclassified with a new number. ***
Quantity
175 units
Distribution
Worldwide Distribution - USA - Australia, Brazil, Cameroon, China, Colombia, Cyprus, The Czech Republic, Finland, France, Germany, Greece, Italy, Japan, Korea, Malaysia, The Netherlands, New Zealand, Norway, Russian Federation, Spain, Sweden, Taiwan, Turkey, UAE, UK, and Canada.
Lot, serial or model codes
Serial numbers: 2791PU2, 2797PU9, 2817PU5, 2824PU1, 2827PU4, 2899PU3, 2900PU9, 2901PU7, 2918PU1, 2919PU9, 2922PU3, 2939PU7, 3095PU7, 3103PU9, 3107PU0, 3219PU3, 3223PU5, 3224PU3, 3230PU0, 3233PU4, 3236PU7, 3358PU9, 3368PU8, 3500PU6, 3513PU9, 3514PU6, 3516PU2, 3523PU8, 3562PU6, 3580PU8, 3957PU8, 3973PU5, 3978PU4, 3994PU1, 4012PU1, 4018PU8, 4025PU3, 4029PU5, 4041PU0, 4043PU6, 4378PU6, 4394PU3, 4397PU6, 4446PU1, 4458PU6, 4767PU0, 4777PU9, 4782PU9, 4787PU8, 4793PU6, 4795PU1, 4797PU7, 4799PU3, 4800PU9, 4803PU3, 4806PU6, 4812PU4, 4813PU2, 4815PU7, 4818PU1, 4822PU3, 4827PU2, 4829PU8, 4830PU6, 4831PU4, 4835PU5, 4839PU7, 4840PU5, 4843PU9, 4844PU7, 4846PU2, 4847PU0, 4852PU0, 4853PU8, 5167PU2, 5169PU8, 5177PU1, 5179PU7, 5182PU1, 5188PU8, 5205PU0, 5208PU4, 5209PU2, 5211PU8, 5213PU4, 5216PU7, 5217PU5, 5219PU1, 5222PU5, 5224PU1, 5225PU8, 5227PU4, 5231PU6, 5233PU2, 5235PU7, 5236PU5, 5237PU3, 5239PU9, 5242PU3, 5244PU9, 5245PU6, 5246PU4, 5247PU2, 5249PU8, 5251PU4, 5260PU5, 5600PU2, 5602PU8, 5608PU5, 5610PU1, 5613PU5, 5621PU8, 5628PU3, 5643PU2, 5648PU1, 5664PU8, 5682PU0, 5683PU8, 5684PU6, 5687PU9, 5691PU1, 6002PU0, 6034PU3, 6036PU8, 6042PU6, 6046PU7, 6051PU7, 6060PU8, 6063PU2, 6093PU9, 6094PU7, 6410PU5, 6424PU6, 6438PU6, 6456PU8, 6474PU1, 6484PU0, 6495PU6, 6498PU0, 6941PU9, 6955PU9, 6959PU1, 6964PU1, 6973PU2, 6986PU4, 6994PU8, 6999PU7, 7307PU2, 7308PU0, 7315PU5, 7317PU1, 7318PU9, 7329PU6, 7330PU4, 7332PU0, 7336PU1, 7338PU7, 7373PU4, 7378PU3, 7379PU1, 7381PU7, 7395PU7, 7396PU5, 7399PU9, 7700PU8, 7719PU8, 7730PU5, 7739PU6, 7754PU5, FIB0033, FIB0046, FIB0048, FIB0051, FIB0058, FIB0065.
510(k) numbers
["K003837"]
PMA numbers
not on file
Product code
JAA
Firm FEI
1720753
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register