Abbott Laboratories, Inc: ACCELERATOR APS System Input/Output Module (IOM); List Number: 7L01-01; manufactured by Inpecco SpA, Segrate (MI), Italy.

Recall number
Z-2221-2009
Recalling firm
Abbott Laboratories, Inc
Product
ACCELERATOR APS System Input/Output Module (IOM); List Number: 7L01-01; manufactured by Inpecco SpA, Segrate (MI), Italy.
Status
Terminated
Initiated
2008-10-27
Posted
2009-09-18
Terminated
2009-12-02
Reason
When an ARCHITECT c16000 System is connected to an ACCELERATOR APS System, the software on the ACCELERATOR APS System may not generate an error warning when a Sample Presentation/Sample Queue error occurs, which could result in contamination of subsequent samples.
Root cause
Software design
Action
Abbott Laboratories notified Customers of the affected device via a "Product Correction" letter dated October 27, 2008. Consignees provided with a temporary work around until a service representative upgrades the system software. For further information, contact your local Abbott Laboratories, Inc. service representative.
Quantity
15 units
Distribution
International Distribution -- Spain, Italy, the United Kingdom, Sweden, Switzerland, the Netherlands, France, Canada and Scotland.
Lot, serial or model codes
Serial Numbers: ABT 042, ABT 055, ABT 037, ABT 016, ABT 041, ABT 046, ABT 035, ABT 040, ABT 039, ABT 033, ABT 028, ABT 011, ABT 006 and ABT 007.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1628664
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register