Skip to the content

FDA Medical Device Recalls (openFDA)

INSTITUT GEORGES LOPEZ: Brand Name: Celsior Cold Storage Solution Product Name: Celsior Cold Storage Solution Model/Catalog Number: 533682 Product Description: In US Celsior is a class II medical device. Celsior is a clear to slightly yellow, sterile, non pyrogenic solution for hypothermic flushing and storage of hearts. Celsior can be used for hypothermic cardiac flushing and storage in preparation for transportation and eventual transplantation of the heart into the recipient. Celsior is packaged in one liter bag with 4 units per carton box. Celsior shall be stored between 36 to 46F. In Europe Celsior is a drug and can be used for thoracic organs (heart & lung) and abdominal organs (liver, kidney, pancreas).

Recall number
Z-2220-2025
Recalling firm
INSTITUT GEORGES LOPEZ
Product
Brand Name: Celsior Cold Storage Solution Product Name: Celsior Cold Storage Solution Model/Catalog Number: 533682 Product Description: In US Celsior is a class II medical device. Celsior is a clear to slightly yellow, sterile, non pyrogenic solution for hypothermic flushing and storage of hearts. Celsior can be used for hypothermic cardiac flushing and storage in preparation for transportation and eventual transplantation of the heart into the recipient. Celsior is packaged in one liter bag with 4 units per carton box. Celsior shall be stored between 36 to 46F. In Europe Celsior is a drug and can be used for thoracic organs (heart & lung) and abdominal organs (liver, kidney, pancreas).
Status
Open, Classified
Initiated
2025-03-13
Posted
2025-07-30
Terminated
not on file
Reason
The potential for risk of microbiological contamination of products due to inability to ensure sterility assurance throughout aseptic manufacturing.
Root cause
Under Investigation by firm
Action
On March 13, 2025 notification emails were sent to OUS customers informing them of the recall and instructing them to check their inventory for affected product and to destroy it. If they did not wish to destroy, to arrange for the product to be returned to the recalling firm. Customers were instructed to acknowledge receipt of the email and provide proof of destruction, if performed.
Quantity
1,532 units
Distribution
International distribution to the country of France.
Lot, serial or model codes
Lot Code: UDI carton box: 23700911900123 & UDI bag: 13700911900126 (for US reference). The product reference recalled in France is different and there is no UDI on this one (not required for medicine in France). Batches concerned by the recall are the following: Lot SL240261-1 Lot SL240266-1 Lot SL240272-1 Lot SL240280-2 These batches have been sold only in France
510(k) numbers
["K991594"]
PMA numbers
not on file
Product code
MSB
Firm FEI
3024820181
Firm city
ST LAURENT DE CHAMOUSSET
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register