Philips Medical Systems North America Co. Phillips: PercuNav Field Generator Stand is a Computed tomography, X-Ray System. Part Number 453561431911. Designed to be used only with the PercuNav" image guided intervention system. The Field Generator Stand (with articulating Mounting Arm) is used to position the Field Generator near the area of interest.

Recall number
Z-2219-2011
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
PercuNav Field Generator Stand is a Computed tomography, X-Ray System. Part Number 453561431911. Designed to be used only with the PercuNav" image guided intervention system. The Field Generator Stand (with articulating Mounting Arm) is used to position the Field Generator near the area of interest.
Status
Terminated
Initiated
2011-04-01
Posted
2011-05-19
Terminated
2012-03-15
Reason
The PercuNav Field Generator Stand could tip over during set up, during use, or during transport.
Root cause
Device Design
Action
On 4/4/11, Philips Healthcare began sending out the URGENT-Medical Device Correction notification letter to their consignees. Philips informed their customers of a potential tipping hazard of the PercuNav Field Generator Stand. The notification letter also provides instructions to customers to ensure the articulating Mounting Arm is positioned so the Field Generator is not further than the specified distance from the back of the clamp on the stand's central pole during set up or during use. Customers are instructed to apply a provided label on stand to warn of a potential for tipping due to excessive force. For transport, customers are instructed to remove the Field Generator from the Mounting Arm and place the Field Generator in the storage compartment on the system cart. The Mounting Arm should then be collapsed and folded into a 90 degree angle. Customers are informed that a Philips field service representative will contact them to schedule installation of a service kit on the stand, free of charge, once service kits are available. Customers can call Philips Customer Service at 1-866-767-7822 and reference to FCO 79500197 for any questions about this recall.
Quantity
65 units total (33 units distributed to US and 32 units to foreign consignees)
Distribution
Worldwide Distribution -- USA, Australia, Austria, Belgium, China, France, Germany, Hong Kong, India, Korea, Netherlands, Singapore, Sweden, Switzerland, Taiwan, Thailand, and United Kingdom.
Lot, serial or model codes
System Code Serial Numbers:  03B8Y3, 20071000005, 20071000006, 20071000007, 20071000008, 20071000009, 20071000010, 20081000011, 20081000013, 20091000019, 20091000020, 20091000021, 20091000022, 20091000023, 02RCHF, 02X3NK, 034K1R, 039VDK, 039W1H, 03B4F3, 03B59R, 03B8MW, 03B8N0, 03B8QR, 03B8XX, 03B8XY, 03B8XZ, 03B8Y0, 03B8Y1, 03B8Y4, 03B8Y5, B0091T, B009BJ, B009HV, B009VJ, B00DDJ, B00DQ2, B00DQX, B00DQZ, B00DRF, B00DRP, B00DT7, B00DXM, B00DY0, B00F1Q, B00F3H, B00F3M, B00HKW, B00HKZ, B00HV4, B00HV7, B00HW3, B00HZR, B00J05, B00P3X, B00P3Y, B00P59, B00P5P, B00PB4, B00PJJ, B00PPP, B00PQ7, and P002A20090500005.
510(k) numbers
["K053610"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register