Ra Medical Systems Inc: The DABRA Laser (model RA-308) is designed to provide ultraviolet laser energy endovascularly through a catheter to occluded lesions in the peripheral arteries to cross them. The system is a self-contained unit that consists of a light source unit and a single use catheter. The light source contains a microprocessor with software to provide safety features and a user-friendly interface. The laser operates by producing a high current discharge within a chamber that is filled with rare gases and halogen. The discharge produces pulses of UV light, which is amplified by reflecting the energy within the resonating chamber many times. The output is directed to the desired target location by the catheter. A footswitch connected to the light source controls the laser action.
- Recall number
- Z-2218-2019
- Recalling firm
- Ra Medical Systems Inc
- Product
- The DABRA Laser (model RA-308) is designed to provide ultraviolet laser energy endovascularly through a catheter to occluded lesions in the peripheral arteries to cross them. The system is a self-contained unit that consists of a light source unit and a single use catheter. The light source contains a microprocessor with software to provide safety features and a user-friendly interface. The laser operates by producing a high current discharge within a chamber that is filled with rare gases and halogen. The discharge produces pulses of UV light, which is amplified by reflecting the energy within the resonating chamber many times. The output is directed to the desired target location by the catheter. A footswitch connected to the light source controls the laser action.
- Status
- Terminated
- Initiated
- 2018-02-15
- Posted
- not on file
- Terminated
- not on file
- Reason
- Lasers/Catheters did not calibrate during set-up prior to use.
- Root cause
- Other
- Action
- On 02/15/18, service technicians started visiting customer facilities to service affected lasers.
- Quantity
- 4
- Distribution
- U.S: CA, PA
- Lot, serial or model codes
- Serial Numbers: 0016, 0039, 0044, 0048
- 510(k) numbers
- ["K170349"]
- PMA numbers
- not on file
- Product code
- PDU
- Firm FEI
- 3004217728
- Firm city
- Carlsbad
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28