Skip to the content

FDA Medical Device Recalls (openFDA)

International Technidyne Corp.: AVOXimeter 1000, AVOXimeter 4000 1000- portable bedside whole blood oximeter that directly measures oxyhemoglobin fraction, the total hemoglobin concentration, and oxygen content. It has enhanced software that optimizes its use int he cardiac catheterization laboratory through calculations of physiologic equations. 4000-a portable bedside whole blood CO-oximeter that directly measures total hemoglobin concentration and the relative concentrations of oxyhemoglobin carboxyhemoglobin concentration methemoglobin. In addition, oxygen content oxygen capacity and oxygen saturation indices are calculated.

Recall number
Z-2218-2011
Recalling firm
International Technidyne Corp.
Product
AVOXimeter 1000, AVOXimeter 4000 1000- portable bedside whole blood oximeter that directly measures oxyhemoglobin fraction, the total hemoglobin concentration, and oxygen content. It has enhanced software that optimizes its use int he cardiac catheterization laboratory through calculations of physiologic equations. 4000-a portable bedside whole blood CO-oximeter that directly measures total hemoglobin concentration and the relative concentrations of oxyhemoglobin carboxyhemoglobin concentration methemoglobin. In addition, oxygen content oxygen capacity and oxygen saturation indices are calculated.
Status
Terminated
Initiated
2011-03-23
Posted
2011-05-19
Terminated
2012-10-25
Reason
An incorrect calibration code that affects the total hemoglobin measurement was applied to affected Avoximeter instruments.
Root cause
Device Design
Action
International Technidyne Corp. sent a letter dated March 23, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory to determine if they have any of the affected product. If they have any of the affected product, customers were instructed to stop using it and remove it from their inventory. Customers were also instructed to complete the attached Customer Account Tracking Form and return it by fax to (732-635-0144) or e-mail (techsupport@itcmed.com) or mail to: ITC Technical Support 20 Corporate Place South Piscataway, New Jersey 08854 For any questions call the Distributor or ITC Nexus Dx Technical Support at 800-631-5945 (US) or 732-548-5700 (International), ext 4707.
Quantity
17
Distribution
Worldwide Distribution - USA including TX, KS, OH, OR, FL, GA, NY, MA, and DC and the countries of Taiwan and Thailand
Lot, serial or model codes
AVOXimeter 1000: K922075 AVOX1000, AVOX1000E-110, AVOX1000INT  AVOXimeter 4000: K951485 AVOX4000, AVOX4000D
510(k) numbers
["K951485"]
PMA numbers
not on file
Product code
DQA
Firm FEI
1000526865
Firm city
Edison
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register