FDA Medical Device Recalls (openFDA)
AngioScore Inc.: AngioSculpt Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter, cutting/scoring, Easy Exchange (EX) delivery system, Manufactured by AngioScore, Fremont, CA. Catheter is for use in the treatment of hemodynamically significant coronary artery stenosis, including in-stent restenosis and complex type C lesions, for improving myocardial perfusion. The devices are available in balloon diameters of 2.0-3.5 mm in 0.5 mm increments, and in scoring balloon lengths of 10 to 15 nm. Catheter length is 139 cm and is compatible with 0.014 inch guide wires and 6F guide catheters. The EX catheter is supplied sterile and intended for single use.
- Recall number
- Z-2218-2010
- Recalling firm
- AngioScore Inc.
- Product
- AngioSculpt Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter, cutting/scoring, Easy Exchange (EX) delivery system, Manufactured by AngioScore, Fremont, CA. Catheter is for use in the treatment of hemodynamically significant coronary artery stenosis, including in-stent restenosis and complex type C lesions, for improving myocardial perfusion. The devices are available in balloon diameters of 2.0-3.5 mm in 0.5 mm increments, and in scoring balloon lengths of 10 to 15 nm. Catheter length is 139 cm and is compatible with 0.014 inch guide wires and 6F guide catheters. The EX catheter is supplied sterile and intended for single use.
- Status
- Terminated
- Initiated
- 2009-12-04
- Posted
- 2010-08-18
- Terminated
- 2011-06-21
- Reason
- Separations of the PTCA catheter shaft at the guide wire exit port have occurred during treatment of coronary artery stenosis. Fragments of the catheter may become lodged in coronary arteries, which resulting in serious patient outcome and require emergency coronary bypass surgery and medical intervention to remove the retained device fragments.
- Root cause
- Process design
- Action
- AngioScore sent an "Urgent: Medical Device Voluntary Recall" dated December 4, 2009 addressed to "Catheterization Lab Manager" at consignees. The Recall notification letter instructed consignees to inspect their inventory to detect product from the involved lot, and segregate it from their inventory. Consignees were asked to record all affected product and fax return forms to the firm. AngioScore Customer Service will contact consignees and provide instructions for return and replacement of affected product.. Consignees can contact AngioScore at 877-264-4692.
- Quantity
- 3870 units
- Distribution
- Nationwide distribution.
- Lot, serial or model codes
- Recall includes only EX Catheter products (all Part/REF Numbers 2034-XXYY) with lot numbers less than (<) F09060003
- 510(k) numbers
- not on file
- PMA numbers
- ["P050018"]
- Product code
- NWX
- Firm FEI
- 3005462046
- Firm city
- Fremont
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28