Skip to the content

FDA Medical Device Recalls (openFDA)

Carestream Health Inc: Carestream Touch Prime, Catalog Number 1738830, and Carestream Touch Prime XE, Catalog Number 1738822 --- Common Name: Touch Ultrasound Diagnostic ultrasound imaging or fluid flow analysis of the human body

Recall number
Z-2217-2016
Recalling firm
Carestream Health Inc
Product
Carestream Touch Prime, Catalog Number 1738830, and Carestream Touch Prime XE, Catalog Number 1738822 --- Common Name: Touch Ultrasound Diagnostic ultrasound imaging or fluid flow analysis of the human body
Status
Terminated
Initiated
2016-07-08
Posted
2016-07-18
Terminated
2017-05-02
Reason
A software issue related to the generic volume measurement functionality could result in the volume of the subject anatomy being overestimated. For example, measurement results of the anatomy may appear enlarged or distended when it actually is not.
Root cause
Software design
Action
The firm, Carestream Health Inc., sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter dated 7/7/2016 and response form the end users/customers on 7/8/2016 via Certified Mail, Return Receipt Requested. The letter described the product, problem and actions to be taken. The Customers were instructed not use the generic volume measurement tool until Carestream has updated the system software. In the interim, calculate volumes by using the specific volume tool from the desired calculation package. If you have distributed the device outside your facility, please alert your customer(s) of this field correction and contact Carestream Customer Care Center as listed above. Carestream has created a software update for Touch Ultrasound to resolve this issue and will update your software. Carestreams representative will contact you and schedule a convenient time to upgrade the software. For questions or concerns, please contact the Carestream Customer Care Center in the US at 1-800-328-2910 (available on a 24/7 basis). Outside of the US, please call your local Service support number.
Quantity
7 Units (US: 5 units, OUS: 2 units)
Distribution
Worldwide Distribution: US Distribution to states of: GA, IA, and TX; and country of: Italy.
Lot, serial or model codes
Touch Prime Serial Numbers: 1037, 1038; -- Touch Prime XE Serial Numbers: 1021, 1079, 1067, 1041, 1043
510(k) numbers
["K152467"]
PMA numbers
not on file
Product code
IYN
Firm FEI
1315356
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register