Reason
In Online Entry (OEx) there are two scenarios where a Test result from one container may file to a Test on another container. In order for this to occur: 1) Site parameter SIT 1,2,1,26 (3,217) [Restrict resulting of tests to owner container (/N)]=Y. 2) The Tests are ordered on the same accession number and 3) The Test codes are defined to have the same upload code as either replicate or
Root cause
Software design
Action
Sunquest sent an Urgent Product Safety Notice dated April 27, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. For the 103 sites that still have this issue Sunquest distribute a new maintenance release. Customers were instructed to use LabAccess Results Workstation (LARS) instead of Online Entry when entering results for instrument method codes that are defined to use replicate tests that are typically ordered on the same accession. Customers were also instructed to share this notification with appropriate laboratory personnel. Customers were questions should call 520-570-2000. For questions regarding this recall call 520-570-2252.
Quantity
103 sites
Distribution
Nationwide Distribution
Lot, serial or model codes
Version 6.3.0 and Later
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1000306472
Firm city
Tucson
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register