FDA Medical Device Recalls (openFDA)

Qiagen GmbH: QIAstat-Dx Respiratory SARS-CoV-2 Panel -IVD intended for the detection and differentiation of nucleic acid from SARS-CoV-2 and the following organism types and subtypes: Adenovirus, Coronavirus 229E, Coronavirus HKU1, Coronavirus NL63, Coronavirus OC43, SARS-CoV-2, Human Metapneumovirus A+B, Influenza A, Influenza A H1, Influenza A H3, Influenza A H1N1/pdm09, Influenza B, Parainfluenza virus 1, Parainfluenza virus 2, Parainfluenza virus 3, Parainfluenza virus 4, Rhinovirus/Enterovirus, Respiratory Syncytial Virus A+B, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae REF 691223

Recall number
Z-2215-2023
Recalling firm
Qiagen GmbH
Product
QIAstat-Dx Respiratory SARS-CoV-2 Panel -IVD intended for the detection and differentiation of nucleic acid from SARS-CoV-2 and the following organism types and subtypes: Adenovirus, Coronavirus 229E, Coronavirus HKU1, Coronavirus NL63, Coronavirus OC43, SARS-CoV-2, Human Metapneumovirus A+B, Influenza A, Influenza A H1, Influenza A H3, Influenza A H1N1/pdm09, Influenza B, Parainfluenza virus 1, Parainfluenza virus 2, Parainfluenza virus 3, Parainfluenza virus 4, Rhinovirus/Enterovirus, Respiratory Syncytial Virus A+B, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae REF 691223
Status
Open, Classified
Initiated
2023-05-09
Posted
2023-07-21
Terminated
not on file
Reason
If cartridges of the affected Serial numbers (SNs) are used, in most cases, the error codes 0xY0003EA and 0xY00045A (where Y is a number between 1 and 4) that lead to a run abortion will occur and no results will be reported
Root cause
Mixed-up of materials/components
Action
Qiagen issued Urgent Medical Device Correction Correction letter via e-mail on 5/9/23. For bounced e-mails, alternative suitable contacts will be identified. Letter states reason for recall, health risk and action to take: Please check if you have remaining stock of LOT 175010712. Do not use the remaining stock of cartridges of LOT 175010712. Dispose of it immediately in accordance to your national and local safety and environmental regulations. Please contact QIAGEN Technical Services for a free-of-charge replacement. Review this notice with your laboratory/medical director. IMPORTANT: Forward this information to all individuals and departments within your organization using the above listed kits. If you are not the end user, please forward this notice to the product end user. Complete the Acknowledgement of Receipt Form attached to this letter by 19. May 2023. Actions taken by QIAGEN QIAGEN was able to identify the root cause of the issue and has implemented immediate actions to ensure that the currently produced cartridges will not cause this issue moving forward. If you have any questions or concerns, please contact your local QIAGEN Technical Services Department through any of the following: Telephone: 800 362 7737 Email: TechService-NA@qiagen.com
Quantity
368 kits
Distribution
AL, AR, CA, DC, FL, GA, ID, KS, KY, MD, MN, NJ, NY,PA, SC, TX, VA
Lot, serial or model codes
GTIN: 14053228038846 LOT Number: 175010712 Affected Serial numbers (SNs) are from 330559516 to 330559783
510(k) numbers
not on file
PMA numbers
not on file
Product code
QJR
Firm FEI
3004013603
Firm city
Hilden
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register