FDA Medical Device Recalls (openFDA)
Medical Components, Inc dba MedComp: 14F x 20CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061423 Product Usage: An aid for introducing a hemodialysis catheter
- Recall number
- Z-2215-2019
- Recalling firm
- Medical Components, Inc dba MedComp
- Product
- 14F x 20CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061423 Product Usage: An aid for introducing a hemodialysis catheter
- Status
- Terminated
- Initiated
- 2019-03-13
- Posted
- not on file
- Terminated
- 2024-03-05
- Reason
- Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
- Root cause
- Labeling Change Control
- Action
- On March 13, 2019, the firm notified customers via Urgent Medical Device Recall letter. Customers were informed that the kits contain a 15F "non-valved" peelable introducer despite the label indicating that the kit is packaged with a "valved" introducer. Customers were asked to do the following: 1. Immediately examine your inventory and quarantine product subject to recall. 2. If you have further distributed this product, please identify your customers and immediately notify them of the recall. 3. Contact your Medical Components, Inc. customer service representative for a Returned Goods Authorization (RGA) number if you have product to return. The phone number to call is 215-256-4201. 4.Complete and return the Medical Device Recall Return Response Form. The form can be returned by fax (215-256-9191) or email (recalls@medcompnet.com). Contact Information for questions: Monday through Friday, 8:00 AM to 4:00 PM, Eastern Standard Time. Phone: 215-256-4201; Email: recalls@medcompnet.com
- Quantity
- 6830
- Distribution
- US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
- Lot, serial or model codes
- MCAK580 MCAS990 MCAY850 MCBA850 MCBH150 MCBN670 MCBY930 MCCC770 MCCK640 MCCN740 MCCR330 MCCW790 MCDC600 MCDK990 MCDR410 MCFB340 MCFG840 MCFL120 MCFP220 MCFT420 MCFZ280 MKAF340 MKAG480 MKAN280 MKAT110 MKAX330 MKBA730 MKBH920 MKBN710 MKBX790 MMAB370 MMAD450 MMAQ940
- 510(k) numbers
- ["K893439"]
- PMA numbers
- not on file
- Product code
- LFJ
- Firm FEI
- 2518902
- Firm city
- Harleysville
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28