Skip to the content

FDA Medical Device Recalls (openFDA)

Medical Components, Inc dba MedComp: 14F x 20CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061423 Product Usage: An aid for introducing a hemodialysis catheter

Recall number
Z-2215-2019
Recalling firm
Medical Components, Inc dba MedComp
Product
14F x 20CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061423 Product Usage: An aid for introducing a hemodialysis catheter
Status
Terminated
Initiated
2019-03-13
Posted
not on file
Terminated
2024-03-05
Reason
Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
Root cause
Labeling Change Control
Action
On March 13, 2019, the firm notified customers via Urgent Medical Device Recall letter. Customers were informed that the kits contain a 15F "non-valved" peelable introducer despite the label indicating that the kit is packaged with a "valved" introducer. Customers were asked to do the following: 1. Immediately examine your inventory and quarantine product subject to recall. 2. If you have further distributed this product, please identify your customers and immediately notify them of the recall. 3. Contact your Medical Components, Inc. customer service representative for a Returned Goods Authorization (RGA) number if you have product to return. The phone number to call is 215-256-4201. 4.Complete and return the Medical Device Recall Return Response Form. The form can be returned by fax (215-256-9191) or email (recalls@medcompnet.com). Contact Information for questions: Monday through Friday, 8:00 AM to 4:00 PM, Eastern Standard Time. Phone: 215-256-4201; Email: recalls@medcompnet.com
Quantity
6830
Distribution
US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
Lot, serial or model codes
MCAK580 MCAS990 MCAY850 MCBA850 MCBH150 MCBN670 MCBY930 MCCC770 MCCK640 MCCN740 MCCR330 MCCW790 MCDC600 MCDK990 MCDR410 MCFB340 MCFG840 MCFL120 MCFP220 MCFT420 MCFZ280 MKAF340 MKAG480 MKAN280 MKAT110 MKAX330 MKBA730 MKBH920 MKBN710 MKBX790 MMAB370 MMAD450 MMAQ940
510(k) numbers
["K893439"]
PMA numbers
not on file
Product code
LFJ
Firm FEI
2518902
Firm city
Harleysville
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register