Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Moore Hip Prosthesis Long Fenestrated Stem 38 mm (1-1/2 Inch) Head Diameter 203 mm Length. Product Usage: The Moore Hip Prosthesis is a femoral him stem that can be used where there is roentgenogrpahic evidence of a satisfactory actabulum and sufficient bone in the femoral neck to seat the prosthesis with or without bone cement.

Recall number
Z-2215-2014
Recalling firm
Zimmer, Inc.
Product
Moore Hip Prosthesis Long Fenestrated Stem 38 mm (1-1/2 Inch) Head Diameter 203 mm Length. Product Usage: The Moore Hip Prosthesis is a femoral him stem that can be used where there is roentgenogrpahic evidence of a satisfactory actabulum and sufficient bone in the femoral neck to seat the prosthesis with or without bone cement.
Status
Terminated
Initiated
2014-07-01
Posted
2014-08-13
Terminated
2015-01-09
Reason
Zimmer is initiating a voluntary recall of 20 production lots of Moore Hip Prosthesis Long Fenestrated Stems due to packaging design verification test failures. Specifically, 9 of 16 test samples failed simulated distribution and shipping testing when the Moore hip stem breached either the inner or outer package tray. Zimmer has not received any complaints of damaged package trays for these 20 lot
Root cause
Packaging process control
Action
On July 1, 2014 Zimmer, Inc. issued recall notification to their consignees via letter. This notice was issued to affected distributors and hospitals with instructions for returning the affected product. Distributors were sent a letter identifying the issue and their responsibilities. These responsibilities include locating and removing the affected product in their territory, as well as identifying hospitals and surgeons who have previously implanted the affected product and providing related contact information (name, address, and phone number). Distributors are to return on-hand affected products to Zimmer and ensure all of their products are accounted for using the form provided in the letter. Hospital risk managers and surgeons were provided with a letter identifying the issue and their responsibilities. These responsibilities include locating the affected product and returning it to their Zimmer sales representative.
Quantity
68
Distribution
Worldwide Distribution - US (nationwide) in the states of IN, WI, MN, OR, and TX, and one foreign consignee in Canada.
Lot, serial or model codes
Part 00402100838, Lot 62435853; Part 00402100841, Lot 62435854; Part 00402100843, Lot 62396506; Part 00402100844, Lot 62424468; Part 00402100848, Lot 62435852; Part 00402100848, Lot 62467218; Part 00402100848, Lot 62511381; Part 00402100851, 37108214; Part 00402100852, Lot 62511377; Part 00402100854, Lot 62511379; Part 00402101241, Lot 62352570; Part 00402101244, Lot 62424469; Part 00402101246, Lot 62400669; Part 00402101248, Lot 62400670; Part 00402101249, Lot 62467222; Part 00402101251, Lot 62470576; Part 00402101252, Lot 62435851; Part 00402101254, Lot 62424470; Part 00402101257, Lot 62524114; & Part 00402101260, Lot 62524115.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JDG
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register