Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: EnCor Breast Biopsy Probes, packaged individually in sterile pouches, 5 pouches/carton, single use, Rx, catalog/REF numbers 12GV Upright (Vertical) - REF ECP0112GV. Product Usage: Indicated for use in acquiring tissue for diagnosing breast abnormalities.

Recall number
Z-2212-2019
Recalling firm
Bard Peripheral Vascular Inc
Product
EnCor Breast Biopsy Probes, packaged individually in sterile pouches, 5 pouches/carton, single use, Rx, catalog/REF numbers 12GV Upright (Vertical) - REF ECP0112GV. Product Usage: Indicated for use in acquiring tissue for diagnosing breast abnormalities.
Status
Terminated
Initiated
2019-05-02
Posted
not on file
Terminated
2022-03-03
Reason
There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product.
Root cause
Component change control
Action
The recalling firm issued letters dated 4/19/2019 via FedEx on 4/19/2019 with proof of delivery to the U.S. customers. The letters sent to foreign countries vary depending on the format required for the country.
Quantity
239 units
Distribution
Worldwide distribution - US Nationwide distribution, including Puerto Rico. There was government and foreign distribution and no military distribution. Foreign distribution was made to Canada, Algeria, Australia, Austria, Belgium, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, India, Ireland, Italy, Japan, Korea, Kuwait, Luxembourg, Malaysia, Martinique, Monaco, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Viet Nam.
Lot, serial or model codes
Lot number VTCW0467.
510(k) numbers
["K040842","K051158"]
PMA numbers
not on file
Product code
KNW
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register